FilmArray Pneumonia Panel (Pneumo), REF: RFIT-ASY-0145, For… recallZ-2440-2023 · BioFire Diagnostics, LLC · initiated May 8, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to manufacturing issue, panels may result in false negative results.
FDA root cause category: Process design
What the company is doing
Between the dates of May 8, 2023 - May 11, 2023, Biomerieux ( Parent company of BioFire Diagnostics, LLC) issued an "Urgent: Medical Device Correction" notification to affected consignees via FedEx. Biomerieux asked consignees to take the following actions: 1. Immediately examine your inventory for product identified in this recall (refer to Table 1). 2. Discontinue use and discard any remaining product in your possession that is subject to this recall. bioMerieux will replace the product at no charge in accordance with BIOFIRE s standard limited warranty. 3. Please complete the accompanying Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to bioMerieux so that bioMerieux may acknowledge your receipt of this notification. 4. If you have further distributed this product, please identify any recipients, and notify them at once of this product recall. 5. For product replacement, as well as any other questions or concerns, please contact our customer support department at [email protected] or via telephone by dialing +1.800.736.6354 and selecting option 5 for Product Technical Support. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report online. " Regular Mail or Fax: Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178.
Recall details
- Recall number
- Z-2440-2023
- Status
- Open, Classified
- Initiated by the company
- May 8, 2023
- Posted by the FDA
- Aug 21, 2023
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- 24 Kits (30 pouches/kit)
- Product code
- QBH
- FDA submissions
- K212727, K180966
- Distribution
- US: CA CO CT FL GA IA IL IN LA MA MI MO MT NC NE NJ NM NY OH OK PA TN UT WV OUS: Canada, India, Japan, Singapore, Colombia
Product description
FilmArray Pneumonia Panel (Pneumo), REF: RFIT-ASY-0145, For FILMARRAY systems, CE 2797, IVD, Rx Only
Code information (models, lots, serial numbers)
Pouch Lot # 2NBT22 /Kit # 2187822 ;UDI-DI: UDI: 00815381020482
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BioFire® Respiratory Panel 2.1 EUA 423738Biofire Diagnostics, LLC · Reagents, 2019-Novel Coronavirus Nucleic Acid3 recalls0 reports
Other devices under the same FDA submission
2 devices cleared under K212727, K180966 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0144Biofire Diagnostics, LLC · Respiratory Panel5 recalls7 reports
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0145Biofire Diagnostics, LLC · Respiratory Panel2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92397).
- Z-2439-2023 · Open, ClassifiedBioFire Respiratory Panel 2.1 (RP2.1), REF: 423742, For FILMARRAY systems, IVD, Rx Only
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.