BETTA LINK SR REUSABLE PRONGED GUIDE- Intended for use in… recallZ-2062-2023 · T.A.G. Medical Products Corporation, Ltd. · initiated May 4, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from T.A.G. Medical Products Corporation, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients includes potential of injury.
FDA root cause category: Device Design
What the company is doing
Zimmer Biomet issued Alert - Field Safety Notice with TAG Letter to consignees by email. on 5/4/23. Letter states reason for recall, health risk and action to take: 1. Immediately discontinue use, sale and distribution of the above products. 2. Please contact Zimmer Biomet at [email protected] for questions and clarifications. 3. Our product surveillance specialists are available to answer questions regarding credit for affected devices in your possession. 4. The attached form must be used and sent back immediately. " Complete the Customer Response Form and return it to [email protected] " Immediately quarantine and return the impacted products to Zimmer Biomet at: Zimmer Biomet Product Service Department ATTN: RECALLS 1777 West Center Street Warsaw, IN 46580 You may direct any questions regarding the recall to [email protected].
Recall details
- Recall number
- Z-2062-2023
- Status
- Open, Classified
- Initiated by the company
- May 4, 2023
- Posted by the FDA
- Jul 3, 2023
- Recalling firm
- T.A.G. Medical Products Corporation, Ltd.
- Product code
- LXH
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Japan, Netherlands.
Product description
BETTA LINK SR REUSABLE PRONGED GUIDE- Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surgeries. Part Number: 110045150
Code information (models, lots, serial numbers)
UDI-DI:10818674025802 Lot Numbers: 22A01, 22C01, 22C02, 22F01
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Betta Link SR Reusable Pronged Guide 110045150T.A.G Medical Products Agricultural Cooperative Association · Orthopedic Manual Surgical Instrument1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92410).
- Z-2060-2023 · Open, ClassifiedBETTA LINK SR KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surger…
- Z-2061-2023 · Open, ClassifiedBETTA LINK LG KNOTLESS IMPLANT KIT-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction surger…
- Z-2063-2023 · Open, ClassifiedBETTA LINK SR REUSABLE FISHMOUTH GUIDE-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction su…
- Z-2064-2023 · Open, ClassifiedBetta Link LG Reusable ProngED Guide - Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction su…
- Z-2065-2023 · Open, ClassifiedBetta Link LG Reusable Fishmouth Guide-Intended for use in soft tissue to bone fixation in the repair of the natural ligament or tendon disruption or to assist in reconstruction su…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page