Masimo LNOP TC-I SpO2, Reusable Ear Sensor, REF: 1794 recallZ-1984-2023 · Masimo Corporation · initiated Apr 21, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Masimo Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
TC-I tip clip sensor may produce readings outside the accuracy specification.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 5/17/2023, Masimo issued a Urgent: Medical Device Recall Notice via FedEx. Masimo asked consignees to take the following actions: 1. Contact Masimo Technical Services to request return of the affected part and lot number(s) for replacements and/ or if you have part numbers or lot numbers not included in this communication and have questions about those parts. 2. Masimo s Technical Services is available Monday through Friday 6:00am -5:00pm PST for live assistance or 24 hours a day to leave a message and someone will contact you. Local contact information can be found at http://service.masimo.com. 3. Complete Attachment 1, then email it to [email protected], or Fax it to 1-949-297-7575 by MM DD, 2023. 4. Any adverse reaction or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program online, by regular mail or by fax.
Recall details
- Recall number
- Z-1984-2023
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2023
- Posted by the FDA
- Jun 21, 2023
- Recalling firm
- Masimo Corporation
- Quantity
- 86 units
- Product code
- DQA
- FDA submissions
- K033296
- Distribution
- US: AK , AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: Argentina, Australia, Austria, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Gibraltar, Guernsey, Hong Kong, Hungary, Iceland, India, Israel, Italy, Japan, Jersey, Jordan, Kenya, Lebanon, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Poland, Portugal, Puerto Rico, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Product description
Masimo LNOP TC-I SpO2, Reusable Ear Sensor, REF: 1794
Code information (models, lots, serial numbers)
UDI-DI: (01)00843997000239 Lot #: K16ANC, K16MWR, K17AEU, 17EGQ
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MasimoSET® LNOP® TC-I SpO2 Reusable Ear Sensor 1794Masimo Corporation · Oximeter, ear1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92371).
- Z-1983-2023 · Open, ClassifiedMasimo RD Set TC-I SpO2 Adult Reusable Ear Sensor, REF: 4053
- Z-1985-2023 · Open, ClassifiedMasimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 1895
- Z-1986-2023 · Open, ClassifiedMasimo LNCS TC-I SpO2, Reusable Ear Sensor, REF: 2503
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page