Oxylog 3000 Plus emergency and transport ventilator recallZ-2018-2023 · Draeger Medical, Inc. · initiated Jun 12, 2023
Class I Open, Classified All recalls posted in July 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.
FDA root cause category: Device Design
What the company is doing
The firm initiated recall notification of customers by mail on 06/12/2023. The Urgent Medical Device Recall Notification indicates that customers should make sure that the device's battery is always removed and reinserted/replaced after an occurrence of the "No int. battery charging, " alarm message without removing the device from the main's supply. Prior to the device being used on battery power, the device should be disconnected and reconnected to an AC main supply and verify the color indicators per the provided diagram. Devices may be used safely as long as the provided instructions are followed. Customers are to ensure all users and maintenance personnel of affected devices are made aware of this recall. Customers are to return the provided Customer Acknowledgement Card utilizing the printed QR code. A Draeger Service representative will contact customers to arrange a firmware update of the Printed Board Assembly Charger. Any questions regarding the operation of your device are to be made to Service Technical Support at 1-800-437-2437 (press 2, 2, 1). Any questions regarding this recall event are to be directed to Michael Kelhart at 267-664-1131 or [email protected]. An updated letter was sent to customers on 02/18/2025 alerting consignees that additional reports of stopped ventilation of devices already updated in which the given alarm No int. battery charging was not handled according to the IFU. As a result of these reports, it was decided to issue an Instructions for Use (IFU) supplement with the check described in the Urgent Medical Device Recall Notice dated June 2023. This IFU was provided with the letter.
Recall details
- Recall number
- Z-2018-2023
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Jun 12, 2023
- Posted by the FDA
- Jul 8, 2023
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 300 units
- Product code
- CBK
- FDA submissions
- K103625
- Distribution
- Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra, Angola, Antigua/Barbados, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bolivia, Brunei Dar-es-S, Bulgaria, Burkina-Faso, Cambodia, Canada, Cayman Islands, Chile, China, Christmas Island, Columbia, Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Faeroe, Finland, France, French Polynesia, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Macau, Malawi, Malaysia, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, North Korea, Norway, Oman, Pakistan, Panama, Pap. New Guinea, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Rwanda, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Sint Maarten, Slovenia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad/Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, Utd. Arab. Emirates, Uzbekistan, Vietnam, Yemen, Zambia, and Zimbabwe.
Product description
Oxylog 3000 Plus emergency and transport ventilator
Code information (models, lots, serial numbers)
Catalog No. 5704811 & 5704813; UDI-DI: 04048675398295; All Serial No.
Other devices under the same FDA submission
1 device cleared under K103625 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Oxylog 3000plus 5704833Drägerwerk AG & Co. KGaA · Ventilator, Continuous, Facility Use4 deaths58 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.