Skip to content

VERIGENE Clostridium difficile Nucleic Acid Test, REF… recallZ-2024-2023 · Luminex Corporation · initiated Jun 5, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Luminex Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.

FDA root cause category: Other

What the company is doing

Luminex issued an Urgent Field Safety Notice - Medical Device Recall notice on 06/05/2023 by email. The notice explained the problem with the stool prep kit, the risk, and requested isolation/quarantine, cease of use, and disposal of the affected products.

Recall details

Recall number
Z-2024-2023
Status
Open, Classified
Initiated by the company
Jun 5, 2023
Posted by the FDA
Jun 28, 2023
Recalling firm
Luminex Corporation
Quantity
114 units
Product code
OZN
FDA submissions
K123197
Distribution
Worldwide and US Nationwide Distribution

Product description

VERIGENE Clostridium difficile Nucleic Acid Test, REF 30-002-022

Code information (models, lots, serial numbers)

UDI/DI , Lot Numbers: 071222022D, 100622022D

Other devices under the same FDA submission

1 device cleared under K123197 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92451).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page