Giraffe OmniBed Carestation CS1 recallZ-2100-2023 · Datex--Ohmeda, Inc. · initiated May 19, 2023

Class II Open, Classified All recalls posted in July 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datex--Ohmeda, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices

FDA root cause category: Nonconforming Material/Component

What the company is doing

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees beginning on 05/19/2023 by mail. The notice explained the problem, risk, and requeested the following: Actions to be taken by Customer/User: Inspect the device bedside panels; if the device has the correct latches, continue to use the device. If the device does not have the correct latches, quarantine the device. Inspect any field replaceable bedside panel stock and quarantine for the affected product. GE Healthcare will replace affected latches.

Recall details

Recall number
Z-2100-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
May 19, 2023
Posted by the FDA
Jul 6, 2023
Recalling firm
Datex--Ohmeda, Inc.
Quantity
4,540 units
Product code
FMZ
FDA submissions
K152814, K213551
Distribution
Worldwide distribution.

Product description

Giraffe OmniBed Carestation CS1, Model Numbers: a. 2082844-001-XXX, b. 2074920-001-XXXXXX, c. M1050205, d. M1135822, e. M1145475, f. M1145476, g. M1235957, h. M1149051, i. M1146098; Labeled as the following: GIRAFFE OMNIBED CARESTATION CS1, GIRAFFE OMNIBED CARESTATION, GIRAFFE OMNIBED, GIRAFFE OMNIBED CS1, OMNI G 115/230V SERVO2 CARESTATION, ASSY-LAU, OMNIBED CHINA EBASE S HUMID, Manufacturing assembly - Make, OBSOLETE-Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/CE Plug, ASSY-LAU, Omnibed Czech S-PH w/Nist O2 w/Reus Probe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Czech S-PH w/Reusable Probe CE Power Cord, Finished good - Make, Omnibed Czech Servo O2 Nist/Scale/Re-usable Probe/SW w/port hole/Ce Plug, ASSY-LAU, Omnibed Indonesian S-GR W/Reuseable Probe CE Plug, Finished good - Make, ASSY-LAU, Omnibed ENG S-Gr w/REUProbe CE Power Cord, Finished good - Make, ASSY-LAU, Omnibed Portuguese S-PH w/NIST O2w/Reusable Probe CE Power Cord, Finished good - Make. Incubator, Neonatal, Warmer, Infant radiant

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

a. Model Number 2082844-001-XXX; GTIN Numbers: 010084068211686221, 010084068211686221; All Lot/Serial Numbers. b. Model Number 2074920-001-XXXXXX; No GTIN Number; Serial Numbers: HDGR52137, HDGL50099, HDGT60164, HDGT60143, HDGT60142, HDGT60139, HDGT60138, HDGT60137, HDGU60753, HDGT60652, HDGU60069, HDGT60141, HDGU60156, HDGU60120, HDGU60119, HDGT60163, HDGT60703, HDGT60245, HDGT60160, HDGT60459, HDGT60246, HDGT60235, HDGU67645, HDGU60065, HDGU60063, HDGU60061, HDGU63122, HDGU63121, HDGU60062, HDGU57534, HDGU67533, HDGU67534, HDGT60460, HDGU67236, HDGT60249, HDGU60201, HDGU60570, HDGU60620, HDGU60023, HDGU60020, HDGT60612, HDGU60022, HDGU60571, HDGU60686, HDGU60202, HDGV60190, HDGW60244, HDGU67677, HDGU67676, HDGU67675, HDGU67280, HDGU67416, HDGU67415, HDGU67382, HDGU67377, HDGU67369, HDGU67366, HDGU67362, HDGU67289, HDGU67284, HDGU67282, HDGU63165, HDGU63167, HDGU63166, HDGV61084, HDGU67436, HDGU67423, HDGU67422, HDGU67421, HDGU67420, HDGU67419, HDGU67418, HDGU67414, HDGU67413, HDGU67381, HDGU67380, HDGU67379, HDGU67378, HDGU67370, HDGU67368, HDGU67367, HDGU67365, HDGU67363, HDGU67361, HDGU67360, HDGU67359, HDGU67358, HDGU67357, HDGU67304, HDGU67303, HDGU67302, HDGU67301, HDGU67300, HDGU67299, HDGU67298, HDGU67297, HDGU67296, HDGU67295, HDGU67294, HDGU67293, HDGU67292, HDGU67291, HDGU67290, HDGU67288, HDGU67287, HDGU67286, HDGU67285, HDGU67283, HDGU67281, HDGJ51159, HDGT60613, HDGT60614, HDGU60700, HDGU60703. c. Model Number M1050205; No GTIN Number; Serial Numbers: HDGV60148, HDGU60656, HDGU60658, HDGT60481, HDGS50047, HDGV60146. d. Model Number M1135822; No GTIN Number; Serial Numbers: HDGL51737, HDGL51738, HDGL51734, HDGM50736, HDGM50841, HDGL51735. e. Model Number M1145475; No GTIN Number; Serial Numbers: HDGM51813, HDGM52120, HDGM52119, HDGN50474, HDGN50472, HDGN50471, HDGN50469, HDGN50466, HDGN50464, HDGN50848, HDGN50847, HDGN50846, HDGN50845, HDGN50844, HDGQ50049, HDGR53203, HDGS51654, HDGS51653, HDGS51652, HDGS51651, HDGS51650, HDGS51649, HDGS51648, HDGS51647, HDGS51646, HDGS51645, HDGM51179, HDGQ50425, HDGN50465, HDGN50849, HDGN50467, HDGN50470, HDGN50475, HDGR52324, HDGR52323, HDGR52322, HDGR52321, HDGN50468. f. Model Number M1145476; No GTIN Number; Serial Numbers: HDGR52896. g. Model Number M1235957; No GTIN Number; Serial Numbers: HDGT60222. h. Model Number M1149051; No GTIN Number; Serial Numbers: HDGR53018, HDGR53085, HDGR52927, HDGR53086, HDGR53021. i. Model Number M1146098; No GTIN Number; Serial Numbers: HDGS52233, HDGS52234.

Other devices under the same FDA submission

1 device cleared under K152814, K213551 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92454).