Skip to content

TruSignal SpO2 recallZ-2031-2023 · GE Healthcare Finland Oy · initiated May 19, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare Finland Oy or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential reduction of energy reaching patient during defibrillation, potential contact with unintended voltage, or inaccurate measurement.

FDA root cause category: Device Design

What the company is doing

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice on 05/19/2023. The notice explained the safety issues associated with the us of the device and requested the actions be taken: Actions to be taken by Customer/User for Safety Issue # 1 and #2: 1. Use an alternate method for SpO2 monitoring such as TruSignal Sensors not impacted by this field action, or an alternate SpO2 device, if possible 2. If alternate methods are not possible, the affected TruSignal SpO2 Sensors can be used for monitoring if they have not been saturated with fluids 3. If defibrillation is necessary, when the affected TruSignal SpO2 Sensors are being used, please follow the instructions below: I. Remove the affected TruSignal SpO2 Sensor (see Table 1 below) from the patient II. Defibrillate the patient, per hospital protocol III. Reattach the affected TruSignal SpO2 Sensor after defibrillation is no longer needed Actions to be taken by Customer/User for Safety Issue #3: 1. Before using Adult/Pediatric SpO2 Sensors (see Table 1), confirm that the sensor does not contain additional material covering the emitter or detector. 2. If any additional material is present, discard the sensor and select another sensor.

Recall details

Recall number
Z-2031-2023
Status
Open, Classified
Initiated by the company
May 19, 2023
Posted by the FDA
Jul 7, 2023
Recalling firm
GE Healthcare Finland Oy
Quantity
512 devices
Product code
DPZ
FDA submissions
K040831
Distribution
Worldwide

Product description

TruSignal Integrated Ear Sensor with GE Connector REF TS-E2-GE; Oximeter, Ear

Code information (models, lots, serial numbers)

GTIN 00840682103138

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92455).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page