HAMILTON-T1 recallZ-2140-2023 · Hamilton Medical, Inc. · initiated Jun 15, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.

FDA root cause category: Software design

What the company is doing

On 06/15/2023, the firm sent via email an "URGENT Medical Device Correction" Letter to customers informing them that, affected HAMILTON-C2/C3/C1/T1 ventilators which are used for long-term ventilation in neonatal group will switch to the "Ambient State" if the ventilation time is exceeding 91 days. During the "Ambient State" the ventilator will alarm audibly and visually displayed on the screen and would result in no active ventilation being provided to patients. Instructions for Customers: Required user actions if failure occurs: In such a case, perform the following actions steps: 1. Provide alternative ventilation immediately. 2. You must switch off ventilator power to exit the Ambient State . 3. Once patient safety is ensured, it is required to have the ventilator serviced. (After passing the service software the device can be returned to use.) For further information see chapter Ambient State in the respective Operator s Manual: HAMILTON-C1/T1 SW version d2.2.X Section 7.6.2 (pg. 130) HAMILTON-C2 SW version d2.2.X section B.8 (pg. 330) HAMILTON-C3 SW version d2.0.X section 7.7.3 (pg. 123) Mandatory periodic, preventive action: To avoid this malfunction, perform the following: The device needs to be restarted regularly. Note: Switching into standby is not sufficient. It is recommend to perform the restart of the device during the regular exchange of the breathing circuits (usually every 28 days). Preventing the failure from occurring through a software update: For HAMILTON-C1/T1: Update the devices to SW version 3.0.0 or higher. -Until the malfunction is corrected as part of this Medical Device Correction, please continue to follow the instructions in this Medical Device Correction. -Clearly display this correction notice at your site. For Questions: Hamilton Medical AG Technical Support at Tel. +41 58 610 10 20 Email: [email protected] Added 08/30/2023 - Firm notified customers in Puerto Rico on ****

Recall details

Recall number
Z-2140-2023
Status
Open, Classified
Initiated by the company
Jun 15, 2023
Posted by the FDA
Jul 21, 2023
Recalling firm
Hamilton Medical, Inc.
Quantity
2319 systems
Product code
CBK
FDA submissions
K181216
Distribution
U.S.: Alabama Alaska Arizona Arkansas California Colorado Connecticut Delaware Florida Georgia Hawaii Idaho Illinois Indiana Iowa Kansas Kentucky Louisiana Maine Maryland Massachusetts Michigan Minnesota Mississippi Missouri Montana Nebraska Nevada New Hampshire New Jersey New York North Carolina North Dakota Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Vermont Virginia Washington West Virginia Wisconsin Wyoming Added 08/30/2023 - Puerto Rico O.U.S.: Bahamas, Japan and Mexico

Product description

HAMILTON T1 Ventilator REF 161006 1610060 161009 1610090 Is intended to provide positive pressure ventilatory support to adults and pediatrics, and optionally infants and neonates

Code information (models, lots, serial numbers)

Show the full code information (14k characters)

Model: HAMILTON-T1 REF: 161006 1610060 161009 1610090 UDI-DI Code: 07630002801850 07630002813532 07630002806091 07630002813549 Serial Numbers: 3057 3058 3063 3070 3077 3081 3085 3089 3093 3112 3115 3119 3121 3124 3126 3127 3129 3131 3132 3133 3136 3138 3140 3168 3169 3185 3191 3193 3195 3199 3200 3235 3236 3241 3243 3245 3247 3249 3250 3258 3259 3263 3275 3280 3299 3307 3309 3311 3312 3313 3317 3320 3340 3574 3576 3577 3579 3580 3604 3605 3613 3616 3618 3621 3629 3657 3658 3661 3663 3666 3668 3672 3691 3702 3703 3716 3731 3731 3739 3744 3746 3747 3748 3749 3750 3751 3761 3785 3792 3803 3824 3829 3832 3840 3848 3852 3853 3854 3862 3869 3875 3886 3889 3890 3902 3921 3933 3937 3953 3960 3961 3971 3973 3974 3983 3986 3987 3989 3990 4017 4046 4050 4050 4052 4052 4053 4057 4066 4070 4075 4076 4086 4090 4093 4094 4096 4100 4102 4118 4121 4123 4244 4339 4341 4349 4354 4355 4416 4422 4429 4502 4516 4518 4524 4527 4534 4535 4551 4552 4583 4584 4592 4593 4595 4596 4599 4600 4629 4639 4696 4724 4726 4729 4730 4731 4775 4794 4915 4920 4928 4965 4968 4982 4992 4993 4996 5000 5001 5002 5007 5011 5034 5052 5057 5062 5063 5069 5074 5075 5113 5116 5120 5169 5181 5234 5238 5239 5246 5260 5261 5264 5267 5268 5271 5275 5277 5287 5288 5292 5293 5294 5296 5297 5300 5301 5302 5303 5305 5308 5311 5312 5314 5315 5319 5320 5323 5324 5326 5327 5328 5331 5332 5334 5335 5337 5339 5340 5341 5343 5397 5421 5427 5428 5430 5431 5435 5438 5439 5447 5450 5454 5460 5474 5481 5483 5484 5485 5486 5487 5488 5489 5490 5491 5492 5493 5494 5495 5496 5497 5499 5501 5504 5547 5552 5585 5669 5676 5678 5680 5688 5691 5692 5694 5695 5696 5697 5698 5699 5700 5701 5702 5703 5704 5705 5707 5708 5709 5713 5714 5716 5719 5720 5721 5722 5724 5727 5728 5736 5739 5742 5752 5756 5756 5760 5833 5834 5835 5882 5901 5961 5993 5994 5995 5996 5998 5999 6001 6015 6024 6025 6026 6028 6030 6031 6032 6034 6035 6036 6037 6047 6048 6051 6057 6058 6068 6069 6097 6098 6099 6101 6117 6120 6122 6123 6124 6125 6166 6170 6172 6188 6191 6213 6214 6216 6221 6225 6226 6238 6277 6280 6304 6309 6311 6314 6317 6323 6330 6353 6354 6357 6360 6361 6374 6394 6395 6398 6400 6401 6433 6437 6441 6528 6538 6643 6645 6648 6652 6661 6683 6686 6689 6691 6695 6697 6700 6714 6715 6721 6752 6764 6788 6821 6846 6854 6855 6875 6885 6888 6893 6894 6895 6896 6899 6910 6937 6952 6962 6966 6969 6974 6978 6983 6988 6991 6995 6998 7000 7004 7010 7011 7085 7089 7095 7097 7128 7152 7154 7155 7157 7160 7161 7166 7167 7182 7184 7185 7187 7193 7242 7245 7246 7248 7250 7251 7263 7266 7271 7281 7349 7350 7351 7352 7358 7361 7433 7474 7480 7624 7627 7635 7654 7665 7669 7675 7676 7677 7681 7687 7689 7692 7696 7697 7708 7714 7719 7725 7727 7735 7736 7740 7741 7745 7748 7749 7754 7755 7756 7757 7758 7768 7769 7779 7785 7790 7790 7794 7839 7919 7934 7937 7952 7968 7973 7975 7976 7978 7988 7995 8007 8034 8068 8071 8074 8079 8080 8082 8093 8096 8102 8109 8110 8113 8121 8126 8128 8130 8138 8265 8277 8284 8287 8302 8339 8343 8389 8396 8406 8407 8408 8417 8418 8419 8426 8427 8431 8435 8439 8440 8442 8450 8451 8453 8458 8463 8466 8467 8468 8480 8481 8487 8519 8522 8523 8525 8527 8600 8602 8604 8609 8611 8613 8617 8618 8622 8624 8628 8644 8681 8696 8707 8746 8759 8760 8761 8765 8767 8768 8772 8774 8775 8781 8801 8806 8812 8829 8831 8836 8844 8848 8849 8850 8851 8853 8859 8865 8872 8883 9016 9052 9146 9183 9186 9187 9190 9191 9251 9255 9258 9261 9272 9343 9356 9359 9379 9386 9388 9430 9431 9432 9433 9434 9435 9510 9512 9513 9514 9515 9516 9542 9543 9549 9554 9558 9571 9572 9574 9582 9610 9616 9623 9630 9635 9649 9653 9658 9673 9700 9704 9765 9974 10012 10021 10028 10036 10042 10043 10049 10050 10051 10052 10054 10056 10057 10063 10064 10065 10068 10069 10108 10191 10192 10209 10216 10258 10265 10287 10288 10290 10298 10321 10322 10324 10327 10329 10331 10334 10335 10341 10345 10351 10355 10356 10357 10358 10372 10378 10381 10388 10391 10393 10401 10404 10408 10414 10420 10422 10423 10426 10427 10491 10499 10573 10581 10584 10591 10592 10624 10627 10634 10641 10658 10670 10686 10687 10688 10689 10691 10692 10693 10694 10695 10696 10698 10699 10702 10703 10704 10718 10731 10793 10802 10834 10868 10869 10873 10917 10933 10946 10966 10968 10969 10973 11007 11012 11013 11026 11067 11070 11072 11073 11074 11075 11084 11119 11184 11187 11196 11201 11207 11208 11212 11213 11214 11215 11216 11217 11218 11220 11222 11225 11226 11227 11228 11234 11235 11236 11237 11240 11245 11246 11250 11251 11252 11254 11256 11258 11260 11262 11326 11328 11329 11330 11332 11333 11334 11350 11351 11359 11360 11374 11378 11380 11381 11383 11389 11392 11447 11453 11460 11527 11530 11531 11532 11556 11563 11564 11567 11570 11571 11572 11573 11768 11794 11838 11841 11842 11843 11845 11847 11849 11854 12036 12038 12093 12139 12140 12141 12142 12143 12218 12220 12230 12270 12271 12272 12292 12293 12294 12295 12307 12310 12311 12312 12313 12326 12327 12328 12329 12330 12336 12338 12339 12341 12342 12343 12347 12367 12371 12372 12374 12375 12388 12389 12418 12419 12420 12421 12422 12424 12425 12430 12431 12432 12433 12435 12453 12454 12456 12459 12462 12464 12494 12550 12632 12642 12644 12645 12650 12651 12655 12666 12668 12669 12670 12671 12672 12680 12686 12731 12735 12737 12776 12817 12818 12900 12948 12949 12951 12952 12966 12967 12968 12969 12973 12977 12978 12979 12981 12984 13001 13002 13005 13016 13017 13018 13023 13026 13027 13183 13261 13262 13372 13419 13422 13425 13426 13427 13557 13558 13572 13727 13730 13731 13732 13733 13734 13735 13736 13737 13738 13740 13741 13742 13743 13744 13745 13746 13747 13748 13750 13751 13752 13754 13755 13756 13757 13758 13759 13760 13761 13762 13763 13764 13766 13767 13770 13771 13774 13778 13779 13781 13782 13783 13785 13786 13787 13788 13790 13791 13792 13793 13794 13795 13796 13797 13798 13799 13800 13801 13832 13940 13946 13948 13950 13951 13954 13956 13962 13967 13975 14015 14025 14026 14027 14028 14029 14030 14031 14032 14033 14035 14040 14042 14044 14054 14066 14067 14068 14069 14071 14072 14074 14075 14077 14089 14096 14106 14109 14112 14126 14131 14148 14321 14330 14618 14620 14660 14790 14791 14792 14828 14886 14899 14907 14913 14950 14952 14956 14960 14995 15031 15058 15062 15063 15064 15068 15127 15209 15214 15215 15217 15248 15250 15257 15264 15325 15345 15346 15359 15363 15365 15366 15370 15448 15449 15456 15479 15523 15597 15620 15621 15631 15635 15642 15644 15649 15681 15734 15751 15761 15762 15766 15769 15791 15792 15794 15795 15796 15797 15800 15802 15803 15804 15806 15807 15813 15815 15816 15820 15821 15824 15826 15827 15828 15829 15834 15837 15842 15847 15849 15861 15869 15874 15906 15911 15912 15914 15919 15929 15930 16009 16050 16060 16061 16083 16086 16091 16095 16097 16098 16099 16125 16126 16127 16128 16129 16131 16132 16133 16134 16140 16146 16147 16148 16149 16154 16166 16167 16168 16169 16170 16171 16173 16174 16180 16186 16188 16189 16190 16200 16241 16292 16311 16327 16336 16337 16340 16346 16351 16352 16353 16355 16356 16358 16367 16378 16379 16380 16381 16395 16400 16404 16405 16413 16419 16423 16424 16427 16429 16434 16435 16436 16438 16442 16443 16470 16483 16509 16512 16516 16517 16519 16545 16548 16579 16580 16588 16593 16601 16614 16615 16618 16625 16626 16629 16630 16644 16647 16650 16652 16653 16660 16665 16673 16675 16682 16686 16690 16697 16700 16710 16741 16742 16743 16744 16745 16750 16751 16753 16756 16757 16759 16797 16798 16799 16800 16803 16814 16822 16827 16830 16834 16835 16838 16842 16916 16922 16923 16925 16926 16927 16928 16929 16930 16931 16943 16944 16946 16948 16950 16953 16960 17000 17001 17006 17056 17066 17067 17082 17370 17425 17563 17565 17600 17605 17606 17608 17610 17611 17654 17656 17658 17659 17660 17661 17662 17668 17669 17670 17675 17677 17678 17679 17684 17685 17687 17689 17696 17698 17699 17701 17709 17720 17741 17768 17850 17872 17873 17874 17876 17877 17913 17922 17925 17941 17946 17949 17955 17957 17969 17975 17976 17982 17985 17986 17991 17993 17995 18009 18016 18045 18049 18100 18200 18216 18244 18247 18260 18279 18330 18333 18338 18344 18345 18349 18352 18353 18356 18363 18364 18367 18368 18370 18373 18376 18377 18378 18379 18383 18384 18387 18390 18392 18414 18415 18418 18419 18421 18433 18498 18501 18502 18503 18504 18507 18509 18526 18537 18550 18589 18603 18639 18643 18647 18651 18692 18713 18721 18724 18725 18727 18736 18737 18739 18743 18746 18748 18750 18754 18757 18758 18759 18763 18765 18767 18772 18778 18779 18782 18784 18788 18789 18790 18793 18804 18807 18808 18816 18888 18948 18949 18979 18980 18993 19265 19266 19267 19290 19292 19294 19298 19300 19304 19310 19311 19362 19364 19365 19367 19371 19372 19376 19385 19386 19387 19394 19398 19452 19470 19479 19488 19490 19492 19501 19510 19530 19588 19589 19677 19689 19710 19724 19727 19728 19738 19796 19802 19807 19809 19810 19827 19834 19841 19893 19898 19899 19901 19904 19917 19918 19920 19921 19940 19944 19947 19958 19961 19963 19966 19967 19969 19971 19976 19977 19978 19991 20000 20002 20010 20012 20023 20030 20035 20043 20050 20054 20055 20061 20072 20074 20075 20076 20079 20085 20088 20092 20099 20102 20103 20105 20109 20111 20114 20126 20141 20156 20157 20159 20184 20237 20246 20249 20250 20252 20256 20258 20261 20264 20266 20267 20268 20269 20372 20374 20375 20376 20378 20379 20380 20381 20448 20453 20467 20491 20496 20498 20502 20513 20523 20527 20530 20531 20532 20534 20536 20537 20539 20540 20541 20542 20545 20546 20549 20550 20551 20552 20553 20561 20562 20647 20689 20694 20697 20701 20705 20706 20708 20710 20711 20712 20725 20746 20749 20751 20755 20756 20757 20758 20760 20762 20778 20779 20820 20862 20867 20880 20916 20965 20981 20986 21130 21133 21145 21154 21156 21158 21205 21245 21258 21261 21263 21266 21267 21269 21271 21285 21286 21289 21291 21294 21295 21353 21354 21397 21401 21405 21410 21411 21418 21419 21422 21544 21553 21567 21870 21873 21887 21986 21989 21990 21999 22064 22111 22112 22113 22142 22143 22156 22169 22172 22175 22176 22178 22180 22194 22195 22200 22201 22202 22203 22204 22205 22206 22208 22961 22962 22963 22967 23001 23002 23005 23007 23009 23077 23087 23089 23099 23243 23245 23252 23274 23284 23383 23384 23385 23386 23387 23388 23389 23390 23391 23392 23393 23397 23402 23405 23406 23431 23438 23448 23449 23450 23452 23455 23456 23463 23467 23485 23486 23489 23493 23505 23536 23546 23550 23582 23589 23665 23948 24009 24013 24018 24020 24022 24039 24080 24104 24152 24165 24170 24171 24173 24183 24269 24301 24303 24304 24305 24326 24328 24329 24330 24341 24343 24556 24562 24566 24568 24569 24570 24573 24607 24609 24611 24613 24622 24626 24686 24688 24690 24692 24693 24700 24701 24702 24704 24773 24777 24780 24810 25428 25433 25443 25444 25445 25447 25450 25455 25456 25693 500025 500091 500210 500234 500261 500300 500371 500397 500434 500463 500511 500524 500536 500647 500789 500816 500866 500963 500983 501096 501166 501171 501223 501323 501337 501445 501448 501523 501544 501563 501593 501594 501746 501789 501814 501894 501913 501946 501997 502000 502028 502069 502112 502130 502349 502364 502421 502426 502492 502707 502708 502828 502834 502868 502868 502895 502975 502993 503091 503108 503112 503131 503175 503183 503189 503197 503209 503217 503221 503280 503292 503292 503313 503341 503349 503354 503391 503520 503561 503571 503572 503582 503597 503597 503599 503605 503638 503639 503659 503721 503783 503815 503816 503824 503892 503900 503913 503990 504102 504115 504231 504254 504289 504352 504353 504356 504410 504467 504472 504506 504524 504573 504587 504591 504657 504676 504690 504718 504740 504929 504955 504975 504991 505008 505043 505076 505089 505110 505123 505153 505208 505209 505274 505300 505302 505348 505394 505401 505512 505547 505548 505556 505613 505642 505727 505770 505810 505816 505821 505940 505954 505988 506044 506056 506069 506070 506139 506146 506161 506162 506205 506220 506228 506236 506330 506334 506336 506353 506368 506405 506425 506440 506480 506481 506489 506578 506589 506599 506636 506669 506778 506842 506895 506952 507010 507034 507042 507088 507097 507102 507105 507127 507130 507140 507169 507339 507342 507421 507477 507503 507505 507522 507642 507647 507681 507699 507703 507706 507720 507742 507759 507767 507788 507823 507830 507833 507848 507850 507926 507943 507945 507955 508004 508048 508102 508150 508239 508281 508298 508300 508320 508321 508416 508420 508439 508496 508560 508586 508610 508627 508660 508682 508718 508731 508732 508752 508763 508794 508831 508883 508933 508957 508971 509016 509028 509038 509046 509053 509080 509086 509089 509096 509115 509119 509143 509167 509168 509182 509183 509214 509222 509294 509348 509394 509402 509419 509445 509448 509529 509590 509598 509653 509658 509674 509683 509695 509728 509735 509772 509796 509826 509838 509840 509843 509909 509915 509919 509921 509922 509925 509926 509927 509928 509929 509930 509931 509932 509986 510016 510016 510021 510022 510056 510063 510068 510071 510123 510127 510140 510142 510145 511293 511296 511303 511305 511424 511435 511440 511448 511448 511476 511487 511487 511489 511491 511491 511500 511503 511510 511512 511515 511527 511541 511547 511561 511562 511580 511587 511591 511592 511595 511598 511599 511601 511602 511608 511614 511623 511626 511643 511645 511646 511650 511651 511657 511661 511664 511665 511669 511670 511673 511678 511685 511691 511693 511701 511704 511707 511709 511718 511723 511745 511749 511801 512063 512104 512111 512116 512126 512136 512138 512141 512154 512163 512166 512168 512189 512190 512193 512230 512302 512326 512331 512335 512337 512339 512352 512363 512386 512408 512447 512522 512586 Puerto Rico Serial Numbers: 4829 5533 11699 15198 17490 17514 18310 18533 24411 500778 501862 502935 504094 504981 505215 505491 505509 506149 506204 507335 508346 509160 509296 509942 509943 509946 509998

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92482).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page