Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part… recallZ-2437-2023 · Datascope Corp. · initiated Jul 31, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Users were identifying autofill failure conditions on the devices causing pump stops.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION dated 7/31/23 was sent to customers. Autofill Alarms User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. As a result of the investigation into the complaints, Getinge would like to reinforce information that is detailed in the Cardiosave Instructions for Use to help minimize the frequency and impact of these alarms: As per the Cardiosave Operating Instructions in sections 2.1.8, Initiation of Assist and 2.1.9, IAB Fill ; the system performs an autofill under the following three conditions: " Initiation of Therapy " Every two (2) hours while therapy is ongoing " If during transport, the atmospheric pressure changes by a 25 mmHg (~1,000 ft.) rise or 50 mmHg (~2,000 ft.) drop during therapy to keep the balloon pressure acclimated to local conditions If the autofill procedure fails to purge and fill the Pneumatic Module properly, the message Autofill Failure will be displayed, an audible alarm will be activated, and pumping will be suspended until the alarm is cleared. Corrective action can be obtained by pressing the Help Available key. The three high priority alarms are: " Autofill Failure No Helium " Autofill Failure Blood Suspected " Autofill Failure Note: Autofill Failure Blood Suspected is caused by a leak in the IAB which results in blood migration back into the IABP system. The failure of the IAB is outside of the scope of this Field Safety Notification. Autofill Failure No Helium " Verify that the Cardiosave device has sufficient Helium in the system. " If the Cardiosave is in Hybrid mode (console within a hospital cart), verify the external Helium Tank is connected to the cart, is opened, and there is sufficient Helium in the tank to support the IABP therapy. " If the Cardiosave is in Rescue

Recall details

Recall number
Z-2437-2023
Status
Open, Classified
Initiated by the company
Jul 31, 2023
Posted by the FDA
Aug 30, 2023
Recalling firm
Datascope Corp.
Quantity
9175 units
Product code
DSP
FDA submissions
K181122
Distribution
Worldwide - US Nationwide distribution.

Product description

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32, UDI: 10607567111117; Model: 0998-00-0800-33, UDI: 10607567109008; Model: 0998-UC-0800-33, UDI: N/A; Model: 0998-00-0800-34, UDI: 10607567111940; Model: 0998-00-0800-35, UDI: 10607567109107; Model: 0998-00-0800-45, UDI: 10607567108421; Model: 0998-00-0800-52, UDI: 10607567108438; Model: 0998-UC-0800-52, UDI: N/A; Model: 0998-00-0800-53, UDI: 10607567108391; Model: 0998-UC-0800-53, UDI: N/A; Model: 0998-00-0800-55, UDI: 10607567108414; Model: 0998-UC-0800-55, UDI: N/A; Model: 0998-00-0800-65, UDI: 10607567113432; Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449

Devices named in this recall

13 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 92480).