Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part… recallZ-2486-2023 · Datascope Corp. · initiated Jul 31, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Users were experiencing a failure in the IAB Fiber Optic Sensor input on the IABP when inserting the Intra-Aortic Balloon Fiber Optic connector.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/23 was sent to customers. Fiber Optic Damage User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. Fiber optics assists in the rapid acquisition and accuracy of the arterial pressure waveform although it is not required to provide therapy with a fiber optic catheter. As detailed within the IFU and educational guides, the patient s arterial pressure can also be obtained via the catheter s inner lumen by connecting a standard fluid-filled pressure bag system or by connecting an external arterial pressure source (typically a radial or femoral arterial line) to the IABP console. Despite the ability of using fiber optics for pressure monitoring, the IFU continues to advise maintenance of the catheter s inner lumen. This guidance is provided within the IFU and Operating guides to both prevent thrombus formation at the end of the catheter, as well as to maintain the integrity of the inner lumen despite the use of the fiber optic signal for monitoring. This provides an alternative source for guiding the therapy if the fiber optic signal is lost. Using AUTO Mode and the ECG as Trigger If in Semi-AUTO MODE and using the fiber optic signal as the trigger source and the signal is not available, utilizing AUTO mode will automatically select the ECG or other available trigger source to guide therapy. When in AUTO mode should the fiber optic signal be lost, therapy can still be safely delivered using only the ECG. However, should the ECG not be a reliable or accurate signal, and the system seeks the arterial signal to guide therapy, the IABP will not have the information needed to guide therapy, and therapy is stopped. Should no alternative arterial pressure source be available, the hospital
Recall details
- Recall number
- Z-2486-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2023
- Posted by the FDA
- Aug 31, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 9379 units
- Product code
- DSP
- FDA submissions
- K181122
- Distribution
- US Nationwide. Global Distribution.
Product description
Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32,UDI: 10607567111117; Model: 0998-00-0800-33, UDI: 10607567109008; Model: 0998-UC-0800-33, UDI: N/A; Model: 0998-00-0800-34, UDI: 10607567111940; Model: 0998-00-0800-35, UDI: 10607567109107; Model: 0998-00-0800-45, UDI: 10607567108421;Model: 0998-00-0800-52, UDI: 10607567108438; Model: 0998-UC-0800-52, UDI: N/A; Model: 0998-00-0800-53, UDI:10607567108391; Model: 0998-UC-0800-53, UDI: N/A; Model: 0998-00-0800-55, UDI: 10607567108414; Model: 0998-UC-0800-55,UDI: N/A; Model: 0998-00-0800-65, UDI: 10607567113432; Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449
Devices named in this recall
13 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
- CARDIOSAVE HYBRID TYPE I (AU) 0998-00-0800-31Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls232 reports
- CARDIOSAVE HYBRID TYPE D PLUG 0998-00-0800-33Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths230 reports
- CARDIOSAVE HYBRID TYPE J PLUG 0998-00-0800-32Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls2 deaths48 reports
- CARDIOSAVE RESCUE, CHINESE 0998-00-0800-75Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls12 reports
- CARDIOSAVE RESCUE, 3.1 EDITION 0998-00-0800-85Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls1 report
- CARDIOSAVE HYBRID TYPE M PLUG 0998-00-0800-35Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls13 reports
- CARDIOSAVE HYBRID TYPE K PLUG 0998-00-0800-34Datascope Corp. · System, Balloon, Intra-Aortic And Control15 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92489).
- Z-2487-2023 · Open, ClassifiedCardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page