PriMatrix Ag recallZ-2082-2023 · TEI Biosciences, Inc. · initiated May 23, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from TEI Biosciences, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possible out of specification endotoxin test results due to issues with in-process and finished goods endotoxin testing.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm sent recall notifications to customers via email on 5/23/2023 and by mail on 5/24/2023. The recall notification informs customers that Integra is conducting a removal of all Surgimend PRS, Surgimend PRS Meshed, Surgimend, Surgimend MP, Revize, Revize-X, PriMatrix Dermal Repair Scaffold, and PriMatrix AG Antimicrobial Dermal Repair Scaffold devices due to identified issues with in-process and finished goods endotoxin testing that may result in out of specification endotoxin results. Customers are to check their stock for any lot of products listed in the provided Appendix and immediately remove them from service. All customers are asked to complete the provided acknowledgement form, regardless if affected product is on hand, and return it by email to [email protected] or by fax to 888-229-0320. Questions about the acknowledgement form can be directed to 888-656-6375. Customers are to keep a copy of the recall notice for their records. A prepaid shipping label has been provided to all consignees so that return of products may be facilitated. Returns are being accepted for credit, however if a refund is preferred the firm asks that thsi is indicated in the "special instructions" section of the acknowledgement form. If any affected devices have been implanted, customers are to monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol. Customers with any questions can reach out to their Integra Sales Representative or Customer Service at 1-800-654-2873.
Recall details
- Recall number
- Z-2082-2023
- Status
- Open, Classified
- Initiated by the company
- May 23, 2023
- Posted by the FDA
- Sep 30, 2024
- Recalling firm
- TEI Biosciences, Inc.
- Quantity
- 10,159 units
- Product code
- KGN
- FDA submissions
- K100261
- Distribution
- Domestic: Nationwide Distribution.
Product description
PriMatrix Ag Meshed; Dermal Repair Scaffold; Rx Only; Sizes 10x12 cm (2:1), 10x25 cm (2:1), 20x25 cm (1:1), 4x4 cm (2:1), 6x6 cm (2:1), 8x12 cm (2:1), 8x8 cm (2:1).
Code information (models, lots, serial numbers)
Item No. (UDI-DI): 607-105-112 (10381780177685), 607-105-125 (10381780113591), 607-105-225 (10381780113607), 607-105-440 (10381780113614), 607-105-660 (10381780177715), 607-105-812 (10381780177739), 607-105-880 (10381780177722).
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- PriMatrix® Ag 607-105-125Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-105-225Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-105-440Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-105-112Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-105-660Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-105-880Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-105-812Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
Product lines
Other devices under the same FDA submission
13 devices cleared under K100261 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- PriMatrix Ag 607-101-112Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-101-125Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-101-225Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-101-440Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-101-660Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-101-812Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-101-880Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-104-440Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-104-660Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix Ag 607-104-880Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-104-812Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
- PriMatrix® Ag 607-104-125Tei Biosciences Inc. · Dressing, wound, drug1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92481).
- Z-2069-2023 · Open, ClassifiedSurgiMend; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 1.0 5x6 cm, 1.0 6x12 cm, 1.0 10x10 cm, 1.0 10x15 cm, 1.0 10x20 cm, 1.0 16x20 cm, 1.0 13x25 cm, 1.0 4x16 cm…
- Z-2070-2023 · Open, ClassifiedSurgiMend PRS; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 10x15 cm Semi-Oval, 7x17 cm Fenestrated, 10x20 cm Fenestrated, 8x16 cm Semi-Oval, 15x15 cm Semi-Oval, …
- Z-2071-2023 · Open, ClassifiedSurgiMend 1.0; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 18x12 cm Thin and 10x15 cm Thin.
- Z-2072-2023 · Open, ClassifiedSurgiMend 2.0; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 5x6 cm, 6x12 cm, 10x15 cm, 16x20 cm, 13x25 cm, 25x40 cm, 20x30 cm, 20x20 cm, 20x25cm, 10x15cm Semi-Ova…
- Z-2073-2023 · Open, ClassifiedSurgiMend MP; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 10x15 cm, 20x20 cm, 13x25 cm, 16x20 cm, 20x25 cm, 20x30 cm, and 25x40 cm.
- Z-2074-2023 · Open, ClassifiedSurgiMend 3.0; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 5x6 cm, 6x12 cm, 10x15 cm, 16x20 cm, 13x25 cm, 25x40 cm, 20x30 cm, 20x20 cm, 20x25 cm, 1x5 cm, 10x25 c…
- Z-2075-2023 · Open, ClassifiedSurgiMend 4.0; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 5x6 cm, 6x12 cm, 10x15 cm, 13x25 cm, 25x40 cm, 20x30 cm, 10x25 cm, 20x0.5 cm, and 10x25 cm Fenestrated…
- Z-2077-2023 · Open, ClassifiedPriMatrix; Dermal Repair Scaffold; Rx Only; Sizes 0.2x26.5 cm (3 pack), 10x12cm, 10x25 cm, 20x25 cm, 4x4 cm, 6x6 cm, 8x12 cm, and 8x8 cm.
- Z-2078-2023 · Open, ClassifiedPriMatrix Fenestrated; Dermal Repair Scaffold; Rx Only; Sizes 4x4 cm, 6x6 cm, 8x8 cm, 14mm Disc, 18mm Disc, 2x2 cm, and 10x25 cm (Meshed 2:1).
- Z-2079-2023 · Open, ClassifiedPriMatrix Meshed; Dermal Repair Scaffold; Rx Only; Sizes 20x25 cm (1:1), 3x3 cm (2:1), 4x4 cm (2:1), 5x5 cm (2:1), 6x6 cm (2:1), 8x12 cm (2:1), and 8x8 cm (2:1).
- Z-2080-2023 · Open, ClassifiedPriMatrix Ag; Dermal Repair Scaffold; Rx Only; Sizes 10x12 cm, 10x25 cm, 20x25 cm, 4x4 cm, 6x6 cm, 8x12 cm, 8x8 cm.
- Z-2081-2023 · Open, ClassifiedPriMatrix Ag Fenestrated; Dermal Repair Scaffold; Rx Only; Sizes 10x25 cm, 20x25 cm, 4x4 cm, 6x6 cm, 8x12 cm, 8x8 cm.
- Z-2083-2023 · Open, ClassifiedRevize; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 10x10 cm, 4x12 cm, 4x16 cm.
- Z-2084-2023 · Open, ClassifiedRevize-X; Collagen Matrix for Soft Tissue Reconstruction; Rx Only; Sizes 12x6 cm and 10x9 cm.
- Z-2085-2023 · Open, ClassifiedTissueMend; Advanced Soft Tissue Repair Matrix; Rx Only; Sizes 5x6 cm, 6x10 cm, and 3x3 cm.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page