Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part… recallZ-2434-2023 · Datascope Corp. · initiated Jul 31, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.
FDA root cause category: Device Design
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/23 was sent to customers. System Over-Temperature User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra-Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. When a System Over Temperature alarm is triggered, the device must be powered down andallowed to cool to a safe operating internal temperature before therapy can be re-initiated using the same IABP.Should you experience a System Over Temperature of Cardiosave IABP during therapy, perform the following instructions documented in CARDIOSAVE Hybrid and Rescue, Operational Instruction for US and OUS: 1. Turn the IABP OFF by pressing and holding the green IABP Power Button, located on theback panel, for 2 seconds 2. Wait 10 seconds 3. Turn the IABP ON by pressing and releasing the green IABP Power Button If the alarm message persists, switch to another IABP if available and contact your Service representative. If another IABP console is not available for use, alternative means of providing hemodynamic support (vasopressors, inotropes or alternate therapies) may be initiated by a healthcare provider as a temporizing measure. Consideration should be given to having a backup IABP available in the event the IABP doesn t cool sufficiently to restart or the alarm reoccurs. To help prevent such alarms from occurring, ensure that there are no restrictions to the airflow around the Cardiosave device. Restrictions to the vents on the device can significantly increase the internal temperature of the device, leading to such temperature excursions. Additionally, as per CARDIOSAVE Hybrid and Rescue Operational Instruction, it is essential not to operate the Cardiosave device outside of the posted operating ambient ranges. This is to prevent such temperature excursions and ensure the safe and effe
Recall details
- Recall number
- Z-2434-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2023
- Posted by the FDA
- Aug 30, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 47 units (42 US, 5 OUS)
- Product code
- DSP
- FDA submissions
- K181122
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85,
Code information (models, lots, serial numbers)
All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449, Model: 0998-UC-0800-75, UDI: N/A; Model: 0998-UC-0800-83, UDI: N/A; Model: 0998-UC-0800-85, UDI: N/A;
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE RESCUE, CHINESE 0998-00-0800-75Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls12 reports
- CARDIOSAVE RESCUE, 3.1 EDITION 0998-00-0800-85Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 92498).
- Z-2433-2023 · Open, ClassifiedCardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page