Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part… recallZ-2431-2023 · Datascope Corp. · initiated Jul 31, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Datascope Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/23 was sent to customers. Docking/Power Battery Failure User Actions to be taken now: Our records indicate that you may have a Cardiosave Hybrid and/or Cardiosave Rescue Intra- Aortic Balloon Pump (IABP) in your facility. Please examine your inventory immediately to determine if you have any Cardiosave Hybrid and/or Rescue IABPs. The Cardiosave device has two operating modes: Hybrid (docked within the hospital cart) and Rescue (transport). When in Hybrid mode the IABP may operate on both AC power and battery power; and when plugged in (AC power) can charge inserted batteries. When in Rescue (transport) mode, the IABP can operate on battery power, or AC Power with the use of the Transport Power Supply accessory (occupying a battery bay). However, unless the Cardiosave is powered off, the Transport Power Supply will not charge a battery inserted into the alternate battery bay unless the Transport Power Supply is connected directly to AC power. There are two methods where the IABP can be in transport/rescue mode without user intent: 1) The Cardiosave Hospital Cart's release latch has been released and the IABP console is not properly installed in the cart. 2) The IABP console is inserted into the hospital cart, but is not fully latched into the cart In both cases, information is provided to the user that the system is not in Hybrid mode via the Rescue Mode Icon on the bottom Cardiosave screen (top center of the screen). In addition, the battery in use message will be displayed in the informational message area even if the AC power cord is plugged in. Additionally, in both scenarios, the Cardiosave will not receive AC power and if the device is in operation, it will continue to operate on battery. When the Cardiosave is not receiving AC power, the batteries are unable to charge. If the AC Plug Icon is not present and/or the Rescue Icon is present, release the latch on the Hospital Cart located b
Recall details
- Recall number
- Z-2431-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 31, 2023
- Posted by the FDA
- Aug 30, 2023
- Recalling firm
- Datascope Corp.
- Quantity
- 9175 units
- Product code
- DSP
- FDA submissions
- K181122
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32,UDI: 10607567111117; Model: 0998-00-0800-33, UDI: 10607567109008; Model: 0998-UC-0800-33, UDI: N/A; Model: 0998-00-0800-34, UDI: 10607567111940; Model: 0998-00-0800-35, UDI: 10607567109107; Model: 0998-00-0800-45, UDI: 10607567108421;Model: 0998-00-0800-52, UDI: 10607567108438; Model: 0998-UC-0800-52, UDI: N/A; Model: 0998-00-0800-53, UDI:10607567108391; Model: 0998-UC-0800-53, UDI: N/A; Model: 0998-00-0800-55, UDI: 10607567108414; Model: 0998-UC-0800-55,UDI: N/A; Model: 0998-00-0800-65, UDI: 10607567113432; Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449
Devices named in this recall
13 devices in the FDA device registry whose identifier appears in the recall's code information.
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
- CARDIOSAVE HYBRID TYPE I (AU) 0998-00-0800-31Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls232 reports
- CARDIOSAVE HYBRID TYPE D PLUG 0998-00-0800-33Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths230 reports
- CARDIOSAVE HYBRID TYPE J PLUG 0998-00-0800-32Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls2 deaths48 reports
- CARDIOSAVE RESCUE, CHINESE 0998-00-0800-75Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls12 reports
- CARDIOSAVE RESCUE, 3.1 EDITION 0998-00-0800-85Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls1 report
- CARDIOSAVE HYBRID TYPE M PLUG 0998-00-0800-35Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls13 reports
- CARDIOSAVE HYBRID TYPE K PLUG 0998-00-0800-34Datascope Corp. · System, Balloon, Intra-Aortic And Control15 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92501).
- Z-2432-2023 · Open, ClassifiedCardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page