OEC 9900 Systems with 9-inch Image Intensifier recallZ-2277-2023 · GE OEC Medical Systems, Inc · initiated May 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE OEC Medical Systems, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
FDA root cause category: Process control
What the company is doing
On 5/15/23 correction notices were mailed to: Director/Manager of Radiology, Hospital Administrator, Head of Radiology Department, PACS Administrator, Director of IT Department, Head, Biomedical Engineering, who were asked to do the following: You can continue to use your device. When transporting the device, ensure it is always in the Anterior Posterior (AP) position, as instructed in the systems Operator Manual. Use caution when moving system to avoid a large impact force. In addition, customers were asked to complete and return the acknowledgement response form to [email protected] Customers with questions or concerns can contact the firm at 1-800-437-1171
Recall details
- Recall number
- Z-2277-2023
- Status
- Open, Classified
- Initiated by the company
- May 15, 2023
- Posted by the FDA
- Jul 31, 2023
- Recalling firm
- GE OEC Medical Systems, Inc
- Quantity
- 908
- Product code
- OXO
- FDA submissions
- K073543, K082781, K120613, K122234
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
Product description
OEC 9900 Systems with 9-inch Image Intensifier
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
UDI-DI: 00840682114301 Serial Numbers: ES-2700, ESXXXL00081, ESXXXL00104, ES-2873, ESXXXX04525, ESXXXX04526, ESXXXX04527, E9-2573, E9XXXX03024, ESXXXX04467, ES-2768, ESXXXX03911, ESXXXX04425, E9-2510, E9-2646, ES-3208, ES-3209, ES-2934, ES-2940, ESXXXX04391, ES-0662, ES-3216, ES-3217, E9-1266, E9-2330, ESXXXX04441, E9XXXX03364, ES-1631, ES-2835, ES-2846, ES-2987, ES-3179, ESXXXX04377, ES-2692, ES-2693, ES-2847, ES-2679, E9-1695, E9-1716, E9-1719, E9-2183, E9-2186, E9-2327, E9-2328, E9-2451, E9-2517, ES-3110, ESXXXX04365, ESXXXX04369, E9-2286, E9-2571, ESXXXX04472, ES-2647, ESXXXX04436, E9-2189, ES-2755, ESXXXX04475, E9-2645, ESXXXX04449, ES-2667, ESXXXX04435, E9XXXX03408, E9-2187, ES-2868, ESXXXX04394, ES-2981, ES-2942, ES-2965, ESXXXX04505, E9-2246, ES-2988, E9XXXX03336, ES-1548, ES-2948, ESXXXX04462, E9-0975, ES-3080, ESXXXX04061, ESXXXX04497, ESXXXX04465, ESXXXX04468, ESXXXX04489, E9-2562, ES-2953, ESXXXX04464, ESXXXX04488, ESXXXX04484, ES-2808, ES-2689, ES-2848, ES-2995, ESXXXL00057, ES-2849, ES-3090, ES-2836, E9XXXX03484, ESXXXX04447, ES-2978, ESXXXL00083, ESXXXX04530, ES-2640, ES-2865, ESXXXX04457, ES-2894, ES-3091, ES-3076, ES-2788, ESXXXX04460, ESXXXX04406, E9-2282, E9-2586, E9-2841, E9-2842, ESXXXX04376, E9-2498, E9-2305, E9XXXX03371, ES-2634, E9-2641, E9-0259, E9XXXX03456, E9-2655, E9-2663, E9XXXX03347, ESXXXL00067, ES-2786, ES-2947, ES-2973, ESXXXL00066, E9-2561, ES-1782, E9-0409, E9-2666, E9-2490, E9-2577, ES-3207, ES-3642, ESXXXL00080, ESXXXX04498, ESXXXX04502, E9-2584, E9-2644, ES-3086, ESXXXL00090, ESXXXX04495, E9-2402, E9XXXX03412, ES-3078, ES-2646, E9XXXX03366, ESXXXL00105, ESXXXL00093, ES-2669, ES-3182, E9-2662, ES-2950, ES-2966, ESXXXL00078, ES-2653, ES-2903, E9-2515, ESXXXX04402, ESXXXX04466, E9-2389, E9-2391, E9-2409, ES-3084, ESXXXL00084, ESXXXL00092, ESXXXL00100, ES-2864, ES-3210, ESXXXX04421, ES-2960, ESXXXX04440, E9XXXX03424, ES-2949-CH, ES-2699, ES-2633, E9XXXX03435, ESXXXX04511, E9-0950, E9-2196, E9-2839, ESXXXL00137, E9-2324, E9-2405, E9-2657, ES-2767, ES-2968, ES-2681, ES-3140, ES-2695, ESXXXX04483, ES-2837, E9-2559, ESXXXX04446, ES-2764, E9-2578, ESXXXX04418, ESXXXL00097, E9XXXX03414, E9XXXX03420, ES-2902, E9-2596, E9-2503, ES-2910, E9-1519, ES-3299, ES-2789, ES-3138, ES-3195, ES-2642, ES-2770, ES-2904, ESXXXX03863, ES-2961, ES-2639, ES-3082, ES-2982, E9-2650, ESXXXX04431, ESXXXX04474, ES-2807, ES-3064, ES-2993, ESXXXX04423, ES-3066, E9XXXX03430, E9XXXX03431, ES-3342, ES-2853, ES-2684, E9-2207, E9-2303, E9-2493, E9XXXX03403, E9XXXX03427, ES-2708, ES-3344, ESXXXX04368, ESXXXX04419, ES-2854, ES-3092, ES-2992, ES-3065, ESXXXX04478, ESXXXX04479, ES-0613, ES-3296, ES-3343, ES-2983, ES-2628, E9-2280, ESXXXX04379, ESXXXX04417, ES-2690, ES-3196, ES-3401, ESXXXX04444, ES-0306, ES-2939, E9-2659, ESXXXX04434, ESXXXX04378, E9-2660, E9XXXX03409, E9XXXX03418, E9XXXX03025, ES-2935, E9-2587, E9-2184, ES-2676, ES-2809, ES-2943, ES-2959, ES-2661, ESXXXX04422, ES-2958, E9-2347, ES-2678, ES-3131, ES-2980, E9-2502, E9-2522, ES-2916, E9XXXX03350, E9XXXX03354, ES-3077, E9-2593, ESXXXX04429, ES-2951, ESXXXL00058, ESXXXX04438, ESXXXX04531, E9-2653, ES-2664, ES-3111, ES-2855, ES-2874, ES-3166, ES-3192, ESXXXX04473, ES-2908, ES-2971, ESXXXX04506, E9-2325, E9XXXX03353, E9XXXX03370, ES-3189, E9-2311, ES-3062, E9-2197, E9-2834, E9-2836, ESXXXX04459, E9-2667, E9-0731, E9-2778, ES-2763, ESXXXX04529, E9-2678, ESXXXL00103, E9-2411, ES-2941, ES-3132, ES-2957, ESXXXX04517, ESXXXX04450, E9-2182, E9-2410, ES-2906, ES-2975, ES-2663, ESXXXX04437, E9-2665, ES-2641, ES-2688, ES-2745, ES-3211, ESXXXX04499, ESXXXX04367, ES-2790, E9-2837, ESCHXX90026, ESXXXX04426, E9-2452, E9-2453, ES-2762, ES-2944, ES-2945, ES-3187, ESXXXX04430, E9-0056, E9XXXX03496, ES-2771, ES-2662, ES-2677, ESXXXX04507, E9XXXX03163, ESXXXX04452, ES-2602, ES-2413, ES-2933, ES-3290, E9-2209, ES-2765, ESXXXL00071, ESXXXX04455, ESXXXL00059, ESXXXL00098, E9XXXX03348, ES-3073, E9-2585, E9XXXX03413, ES-1067, ES-2769, ESXXXX04445, E9-2591, E9-2656, ES-2824, ES-2905, ESXXXL00091, ESXXXX04469, ESXXXL00079, E9-2509, ESXXXX04403, ESXXXX04480, ESXXXL00102, E9-2664, ES-2986, ESXXXX04405, ES-3240, ES-2656, ES-2671, ESXXXL00074, ESXXXX04513, ES-3165, ES-2670, ES-3099, ESXXXX04461, ESXXXL00099, ESXXXL00070, ES-2850, ESXXXL00072, ES-3169, ESXXXL00060, ESXXXL00086, ESXXXL00088, ESXXXL00077, E9-2495, ES-2346, ESXXXL00082, ESXXXL00094, E9-2518, ES-3224, E9-0935, E9-2575, E9-2832, E9-2337, E9-2534, ESXXXL00106, E9-2535, ES-2654, ES-2851, ES-3133, ES-3298, E9-2581, ESXXXX04049, ES-7227-CMH, E9XXXX03001, ESXXXX04428, ES-2645, ESXXXX04404, E9-2519, E9-2661, E9-2385, E9-2564, ES-2956, ES-2731, ES-2472, ES-2976, E9XXXX03477, ESXXXX04392, ESXXXX04504, ESXXXX04512, ES-2955, ES-3538, ESXXXX04477, ES-3139, E9-2563, E9XXXX03352, E9-2497, E9XXXX03356, E9-0231, E9-2195, ES-3089, ES-3183, ESXXXX04496, ES-2962, ES-2963, ESXXXX04442, ESXXXX04424, ESXXXX04393, E9-2583, ES-2969, E9XXXX03349, ESXXXX04491, ESXXXX04493, ESXXXX04494, ESXXXX04519, ES-1598, ES-3190, ESXXXX04486, ESXXXX04487, E9-2256, E9-2582, ESXXXL00085, ESXXXX04508, E9-2579, ES-2996, E9-2594, ES-2683, ES-2858, E9-2899, ES-2756, ES-2774, ES-2937, ES-2761, ES-2867, ES-2694, ES-2655, ES-3063, ES-2964, ESXXXL00056, ES-2892, ES-2680, ES-3058, ES-3297, E9-2516, E9-2520, ES-2890, ES-2629, ESXXXX04514, ES-3152, ES-3153, ESXXXX04407, ESXXXX04427, ES-2649, ES-2773, ES-2776, ES-3255, ESXXXX04458, ESXXXX04470, E9-2190, ES-2668, ES-3745, E9-2285, E9-2566, ESXXXX04338, E9-2206, E9XXXX03425, ES-2659, E9-2565, ES-2866, ES-2869, E9-2201, E9-2202, E9-2499, E9XXXX03433, E9-2580, ESXXXL00069, E9-2203, E9-2204, ES-2686, ESXXXL00073, ES-2946, E9-2390, E9-2643, E9-2556, E9-2395, E9-2572, ES-3079, ES-3081, E9-2638, ESXXXX04485, E9-2679, ES-3300, ESXXXX04366, E9-2211, E9-2284, E9-2304, E9-2346, E9-2407, E9-2511, E9-2512, E9-2569, E9XXXX03485, ES-0080, ES-2652, ES-2954, ES-2997, ES-3598, ES-2775, E9-2831, ESXXXX04451, ES-2666, ES-2915, ES-3341, ES-3056, ES-1052, E9-2404, ES-2650, ES-3112, ES-3301, ESXXXX04453, ES-2665, E9-2398, E9-2399, E9-2318, E9-2392, ES-2691, ES-3204, ESXXXL00075, ESXXXL00076, ESXXXX04439, E9-2532, ES-2635, ES-3180, ES-2914, ES-2952, ESXXXX04509, ES-2787, ESXXXL00061, E9-2298, E9-2200, E9-2557, E9-2889, ES-2825, ES-2907, ES-3135, ES-3136, ES-3137, ESXXXX04501, E9-2245, E9-2297, ES-2754, E9-2296, E9-2523, ESXXXX04521, ES-3134, ES-2660, ESXXXX04448, ES-2994, ESXXXX04482, ESXXXX04443, E9-2329, E9-2397, E9-2401, E9-2640, E9-2642, E9-2652, E9XXXX03419, E9XXXX03493, ES-2648, ES-2687, ES-2772, ES-2834, ES-2909, ES-2972, ES-3074, ES-3215, ES-7307-CMH, ESXXXX04481, ESXXXX04492, ES-2685, ESXXXX04463, E9-2210, E9-2567, ES-3193, ES-3203, ES-3167, ESXXXL00087, ES-2810, E9XXXX03365, ES-2637, ES-2893, ES-3040, E9-2639, ES-2856, E9-2393, E9-2345, ESXXXL00089, E9XXXX03437, ESXXXX04516, ES-2970, E9-2185, E9XXXX03436, E9-2521, ES-2889, ESXXXX04370, ESXXXX04490, E9-2654, E9-2496, ES-2999, ES-3285, E9-2588, ES-2682, ES-2979, ES-2630, ES-2991, E9-2568, E9-2205, ES-3083, E9-2551, E9-2833, E9-2835, ES-3085, ES-3191, ES-3194, E9-2574, ES-2636, E9-2394, ES-2990, ES-3093, E9-2322, ESXXXX04420, ES-2967, ES-2974, ESXXXL00101, E9XXXX03006, ES-2651, ES-3188, ESXXXL00095, ESXXXL00096, ES-2998, E9XXXX03405, E9XXXX03404, ESCHXX90030, ES-7328-CMH, ES-7336-CMH, ESCHXX90051, ESCHXX90025, ESCHXX90049, E9XXXX03351, ES-2823, E9-2295, E9-2277, E9-2283, E9XXXX03460, E9XXXX03467, ESCMHX07474, ESCMHX07475, ESCHXX90035, ESCHXX90037, E9-2494, ESCHXX90034, ESCHXX90029, E9XXXX03372, ESCHXX90043, ESCHXX90042, E9XXXX03373, ESCHXX90033, ESCHXX90036, E9-2403, ES-2936, ES-3164, ESXXXX04515, ES-2698, E9-2208, E9-2514, ESXXXX04433, E9XXXX03407, E9-2491, ES-3163, ES-3213, ES-3214, E9-2293, E9-2492, E9-2648, E9-2651, E9XXXX03377, ES-3168, E9-2263, ES-7322-CMH, ES-7332-CMH, E9XXXX03423, ES-2938-CH, ES-7331-CMH, ES-7330-CMH, E9XXXX03428, ESCMHX07468, E9-2270, E9-2273, ESCMHX07478, ESCMHX07479, E9-2259, E9-2254, E9XXXX03406, ES-7168-CMH, E9-2252, E9-2271, E9-2272, ES-7323-CMH, ES-7335-CMH, E9-2275, E9-2274, E9XXXX03479, ES-7317-CMH, ESCMHX07465, ESCMHX07464, ES-7337-CMH, E9XXXX03495, ESCMHX07467, ESCMHX07462, ES-7326-CMH, ES-7316-CMH, ESCMHX07477, E9XXXX03468, E9XXXX03472, E9XXXX03473, E9XXXX03363, E9XXXX03469, E9XXXX03440, E9XXXX03476, E9XXXX03478, E9-2396, E9-2400, E9-2408, ES-3072-CH, E9-2570, E9-2255, E9-2253, E9-2290, E9-2288, E9-2289, E9-2294, E9-2406, E9-2501, E9-2748, ES-3069, ESXXXX04523, ESXXXX04524, ES-2985, E9-2488, E9-2261, E9-2508, ES-2891, E9-2506, E9-2500, ES-3075, E9-2507, E9XXXX03434, ES-2984, E9XXXX03426, E9-2589, E9-2838, ES-2852, E9-2560, ES-3450, ES-2870, ESXXXX04503, ES-3038, E9-2258, E9XXXX03005, E9-2262, E9XXXX03474, ESCHXX90039, ESCHXX90040, ESXXXX04380, ESXXXX04454, ESXXXX04476, ESXXXX04518, ESXXXX04520, ES-3070, E9-2533, ES-3068, ESXXXX04528, E9-2531, E9-2590, E9-2747, ES-2911, ES-3071, E9XXXX03475, ESXXXX04471, E9XXXX03359, E9XXXX03368, E9XXXX03491, E9XXXX03471, E9XXXX03415, E9XXXX03439, E9XXXX03357, E9XXXX03429, E9XXXX03470, E9XXXX03488, E9XXXX03490, E9XXXX03492, E9XXXX03494, ESXXXX04500, E9XXXX03487, E9-2317, E9XXXX03367, E9XXXX03416, E9XXXX03438, ES-7346-CMH, E9XXXX03466, E9-2257, ES-3087-CH, ESCHXX90041, E9-2302, ESCHXX90044, ESCHXX90047, ES-7325-CMH, ES-7348-CMH, ES-2977-CH, ES-1717-CH, ES-3088-CH, ESCHXX90045, ESCHXX90048, ESCHXX90046, ESCMHX07476, ESCMHX07466, E9-2279, E9XXXX03421, ESCHXX90053, ESCHXX90028, ESCHXX90052, ESCMHX07473, ESCHXX90032, ESCHXX90027, E9XXXX03459, ES-7324-CMH, E9XXXX03480, E9XXXX03481, ESXXXX04522, E9-2323, E9XXXX03417, ESCMHX07472, E9XXXX03369, E9XXXX03374, E9-2326, E9-2558, E9-2281, E9-2276, E9-2291, E9-2292, E9XXXX03486
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- OEC 9900 ELITEOec Medical Systems, Inc. · Interventional fluoroscopic x-ray system3 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92504).
- Z-2276-2023 · Open, ClassifiedOEC 9800 Systems with 9-inch Image Intensifier
- Z-2278-2023 · Open, ClassifiedOEC Elite Systems with 9-inch Image Intensifier
- Z-2279-2023 · Open, ClassifiedField Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page