OmniTom/OmniTom Elite- X-ray computed tomography… recallZ-2068-2023 · NeuroLogica Corporation · initiated Jun 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from NeuroLogica Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment
FDA root cause category: Process design
What the company is doing
NeuroLogica Corporation issued Urgent Medical Device Correction letter to customers via email June 6, 2023. Letter states reason for recall, health risk and action to take: Users may continue to use the NL5000 system with caution, ensuring they are aware of the potential drive issues due to the wheels. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. NeuroLogica will undertake the voluntary corrective action considering the possibility of delay in patient treatment if the scanner is not operational due to battery failure, as well as the potential for the scanner to halt mid scan. The hardware upgrades provide additional safeguards by improving the design to minimize any risks of the above situations from happening. Service engineers will correct all affected devices free of charge and contact you to arrange for the correction starting June 2023. All corrections in the field will be completed by December 31, 2023. Contact customer service team at 1-888-564-8561 (US/Canada) or 978-564-8561 (International) and [email protected].
Recall details
- Recall number
- Z-2068-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 6, 2023
- Posted by the FDA
- Jul 3, 2023
- Recalling firm
- NeuroLogica Corporation
- Quantity
- 48 units
- Product code
- JAK
- FDA submissions
- K202526, K171183
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, Grand Bahama, India, Italy, Korea, Portugal, Saudi, Scotland, Thailand.
Product description
OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck Model Number: NL5000
Code information (models, lots, serial numbers)
GTIN: 10815411020335 10815411020663 Serial Numbers: 10 14 23 25 26 36 38 40 41 42 44 46 48 49 50 51 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 15R
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- OmniTom Elite 0-NL5000-002Neurologica Corp. · System, X-Ray, Tomography, Computed1 recall17 reports
- CT SCANNER, NL5000 0-NL5000-001Neurologica Corp. · System, X-Ray, Tomography, Computed1 recall12 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page