Perifix recallZ-2002-2023 · B. Braun Medical, Inc. · initiated Jun 5, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B. Braun Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits were assembled with the incorrect Filter Straw.
FDA root cause category: Mixed-up of materials/components
What the company is doing
Customers received a recall notification dated 06/05/2023 by mail notifying them that affected lots of Epidural Kits were inadvertently assembled with the incorrect Filter Straw. The recall notification asks customers to determine if any affected devices are in inventory and subsequently quarantine any identified product, return the Product Recall Acknowledgement Form notating any affected devices currently in quarantine, and return the form by fax to 610-849-1197 or by email to [email protected]. Coordination of product return for provision of credit/replacement will commence upon return of the form. If product has been further distributed the firm asks that the recall notification be forwarded to those consignees down to the consumer level. Customers with any questions can direct them to BBMI's Postmarket Surveillance Department at 1-883-425-1464.
Recall details
- Recall number
- Z-2002-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 5, 2023
- Posted by the FDA
- Jun 23, 2023
- Recalling firm
- B. Braun Medical, Inc.
- Quantity
- 16,070 units
- Product code
- OGE
- Distribution
- US Nationwide distribution.
Product description
Perifix / Epidural anesthesia kit (10 count carton)
Code information (models, lots, serial numbers)
Model No. 332217, 332097, 332220, 332078; GTIN: 04046964178283, 04046964177361, 04046964177361, 04046964178320, 04046964177200; Lot No. 0061850767, 0061855339, 00618553400061858554, 0061837264.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Perifix® 332097B. Braun Medical Inc. · Epidural anesthesia kit1 recall39 reports
- Perifix® 332220B. Braun Medical Inc. · Epidural anesthesia kit1 recall21 reports
- Perifix® 332078B. Braun Medical Inc. · Epidural anesthesia kit1 recall12 reports
- Perifix® 332217B. Braun Medical Inc. · Epidural anesthesia kit1 recall2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92523).
- Z-2003-2023 · Open, ClassifiedDesign Options / Epidural anesthesia kit (10 count carton)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.