Skip to content

Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre… recallZ-1971-2023 · Ellex Medical Pty Ltd. · initiated Mar 16, 2023

Class II Open, Classified All recalls posted in June 2023

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ellex Medical Pty Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Ellex Safety Bulletin is the customer notification letter. Ellex will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. For questions contact your local Ellex-Lumibird representative or email us on [email protected].

Recall details

Recall number
Z-1971-2023
Recall class
Class II
Status
Open, Classified
Initiated by the company
Mar 16, 2023
Posted by the FDA
Jun 27, 2023
Recalling firm
Ellex Medical Pty Ltd.
Quantity
57 systems in total (6 units shipped to United States)
Product code
HQF
FDA submissions
K081565, K142398
Distribution
US nationwide/ Worldwide Distribution

Product description

Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G

Code information (models, lots, serial numbers)

Serial Nos., UDI IPY0374, (01) 0 9342395 00012 0 ISRY0242, (01) 0 9342395 00015 1 ISG0370, (01) 0 9342395 00024 3 IPG0539, (01) 0 9342395 00009 0 BG1190

Other devices under the same FDA submission

2 devices cleared under K081565, K142398 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page