Catalyft LS Expandable Interbody System recallZ-2118-2023 · Medtronic Sofamor Danek USA Inc · initiated Jun 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is the potential for out of specification device geometries. widths, lengths, heights, and lordotic angles to accommodate patient anatomy. These devices can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. Implants have a central cavity that allows them to be packed with autogenous bone graft and/or allograft bone graft comprised of cancellous and/or Corticocancellous bone, and/or demineralized allograft bone with bone marrow aspirate.
FDA root cause category: Under Investigation by firm
What the company is doing
Medtronic issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 06/06/2023 by UPS 2-day delivery. The notice explained the problem , risk, and requested the following actions be taken: "" For purchased inventory not consumed during a procedure: o Identify and quarantine any unused impacted product(s). Refer to the table above for impacted products. o Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form. " For Medtronic owned product at your facility: o Your Medtronic Sales Representative will locate and return this product. " Complete the Customer Confirmation Form enclosed with this letter (even if you have no product to return), acknowledging that you have received this information. " This notice should be distributed to all others in your organization who should be aware, or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records."
Recall details
- Recall number
- Z-2118-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 6, 2023
- Posted by the FDA
- Jul 10, 2023
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 8 units
- Product code
- OVD
- FDA submissions
- K212653
- Distribution
- US Nationwide distribution in the states of IN, WI, GA, OR, AR, OH, TN, TX, NC, OR, NE, MN, FL.
Product description
Medtronic Catalyft LS Expandable Interbody System, Anterior Standalone Cage, REF 981023325, spinal implant
Code information (models, lots, serial numbers)
UDI/DI 00763000465872, Lot Numbers: NM21J004, NM22A032
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Catalyft™ LS Expandable Interbody System 981023325Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
Product lines
Other devices under the same FDA submission
5 devices cleared under K212653 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Catalyft™ LS Expandable Interbody System 981023322Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall2 reports
- Catalyft™ LS Expandable Interbody System 981025522Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall1 report
- Catalyft™ LS Expandable Interbody System 981025525Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Catalyft™ LS Expandable Interbody System 981027722Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Catalyft™ LS Expandable Interbody System 981027725Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92567).
- Z-2117-2023 · Open, ClassifiedMedtronic Catalyft LS Expandable Interbody System, Anterier Standalone Cage, REF 981023322, spinal implant
- Z-2119-2023 · Open, ClassifiedMedtronic Catalyft LS Expandable Interbody System, Anterior Standalone Cage, REF 981025522, spinal implant
- Z-2120-2023 · Open, ClassifiedMedtronic Catalyft LS Expandable Interbody System, Anterior Standalone Cage, REF 981025525, spinal implant
- Z-2121-2023 · Open, ClassifiedMedtronic Catalyft LS Expandable Interbody System, Anterior Standalone Cage, REF 981027722, spinal implant
- Z-2122-2023 · Open, ClassifiedMedtronic Catalyft LS Expandable Interbody System, Anterior Standalone Cage, REF 981027725, spinal implant
- Z-2123-2023 · Open, ClassifiedMedtronic Catalyft LS Expandable Interbody System Implant Kit, Anterior Standalone Cage, REF SPS03188, spinal implant kit
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page