Skip to content

RET Distal Probe. Used in Fluorescence in situ… recallZ-2210-2023 · Cytocell Ltd. · initiated May 24, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cytocell Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

DNAs used during the manufacture of two affected models were accidentally inverted during manufacturing.

FDA root cause category: Process control

What the company is doing

Customers were sent a recall notification dated 05/24/2023 informing them of this recall event. Customers are asked to identify any affected devices in stock and quarantine any devices from recalled lots. Affected devices are to be destroyed by customers. Additionally, customers are encouraged to review results obtained utilized affected devices to ensure that no results were misinterpreted due to the incorrect labeling of the DNA. Customers are to return the provided response form by fax to +44(0)1223294048 or by email to [email protected]. Customers with any questions can contact Cytocell by phone at +44(0)1223294048, by fax at +44(0)1223294986, or by email at [email protected].

Recall details

Recall number
Z-2210-2023
Status
Open, Classified
Initiated by the company
May 24, 2023
Posted by the FDA
Jul 20, 2023
Recalling firm
Cytocell Ltd.
Quantity
61 units
Product code
MVU
Distribution
US Nationwide distribution in the states of CA, FL, OH, & SC.

Product description

RET Distal Probe. Used in Fluorescence in situ hybridization (FISH), a laboratory technique for detecting and locating a specific DNA sequence on a chromosome.

Code information (models, lots, serial numbers)

Catalog No. LPS509-A; UDI-DI: 05055844900516; Lot No. 076764, 080124, 082129, 082560, 083247, 083550, 083850.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 92586).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page