Skip to content

Zimmer Biomet Patient Specific Instruments recallZ-2213-2023 · Materialise N.V. · initiated Apr 26, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Materialise N.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The wrong tibia guide was included intended for a different patient case.

FDA root cause category: Process control

What the company is doing

Materialise notified Zimmer Biomet international logistics on April 26, 2023 via email and telephone requested return of cases. Zimmer BIomet confirmed that they would directly contact the customer to return the original cases ZB23-MAN-OLA and ZB22-UHI-NEF. Materialise shipped replacement cases on May 2, 2023 and May 3, 2023. For questions call +1-734-259-7010 or email: [email protected].

Recall details

Recall number
Z-2213-2023
Status
Completed
Initiated by the company
Apr 26, 2023
Posted by the FDA
Jul 20, 2023
Recalling firm
Materialise N.V.
Quantity
2 Guides
Product code
JWH
FDA submissions
K221337, K173445
Distribution
International Distribution to countries of: Germany, Netherlands

Product description

MRI Biomet Pin Guide for Total Knee Arthroplasty (Zimmer Biomet Patient Specific Instruments (PSI)-- patient-matched device designed to fit the patients anatomy to transfer a patient-specific pre-operative plan to the OR. Ref: KNM-ZB02-00-01

Code information (models, lots, serial numbers)

UDI-DI: (01)05420060310027 Lot Number/Case Number: ZB22UHINEF, ZB23MANOLA

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K221337, K173445 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page