Why it was recalled

Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).

FDA root cause category: Nonconforming Material/Component

What the company is doing

Siemens issued Customer Safety Advisory Notice via AX015/23/S on 6/6/23. Letter states reason for recall, health risk and action to take: Siemens will correct the hardware error via Update Instruction AX012/23/S. This measure will replace the current single tank cover of all affected systems. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein

Recall details

Recall number
Z-2240-2023
Status
Open, Classified
Initiated by the company
Jun 5, 2023
Posted by the FDA
Jul 27, 2023
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
9 units
Product code
OWB
FDA submissions
K210309
Distribution
Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY and the countries of Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg, Mauritius, Mexico, Monaco, Peru, Philippines, Poland, Romania, Slovakia, South Africa, Spain, Switzerland, Taiwan, Ukraine, United Kingdom.

Product description

Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088

Code information (models, lots, serial numbers)

UDI-DI :04056869148243 US S/N: 30001 30002 30008 30013 30029 30031 30032 30033 30034