ARTIS icono floor-Angiography systems developed for single… recallZ-2360-2023 · Siemens Medical Solutions USA, Inc · initiated Jun 19, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
FDA root cause category: Process control
What the company is doing
Siemens issued Urgent: Medical Device Correction to all users of ARTIS icono/ ARTIS pheno systems with a specific Polydoros ACX power generator on 6/19/23. Letter states reason for recall, health risk and action to take: It is strongly recommended to establish appropriate emergency procedures until the corrective action has been performed. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. Siemens will correct the hardware error via Update Instruction AX008/23/S. This measure will correct the Polydoros ACX reactor coil of all affected systems. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
Recall details
- Recall number
- Z-2360-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 19, 2023
- Posted by the FDA
- Aug 8, 2023
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 7 units
- Product code
- OWB
- FDA submissions
- K220432
- Distribution
- Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Latvia, Luxembourg, Macedonia, Mexico, Monaco, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Product description
ARTIS icono floor-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327700
Code information (models, lots, serial numbers)
UDI-DI: 4056869149325 Serial Numbers: 170561 170549 170559 170546 170552 170571 170570
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
Other devices under the same FDA submission
1 device cleared under K220432 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92653).
- Z-2359-2023 · Open, ClassifiedARTIS icono biplane -Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327600
- Z-2361-2023 · Open, ClassifiedARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 10849000
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page