MEDLINE NON-STERILE PVP SOLUTION, packaged in kits as… recallZ-2132-2023 · MEDLINE INDUSTRIES, LP - Northfield · initiated May 18, 2023
Class II Open, Classified All recalls posted in July 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Affected product lots were mislabeled as "STERILE PVP SLN" when they should have been titled "NON-STERILE PVP SOLUTION, STERILE PACKAGING".
FDA root cause category: Process control
What the company is doing
Medline issued a recall notice on 06/02/2023 to its consignees via USPS first class mail. The notice explained the problem with the product and requested the destruction of the product. Medline identified additional products, lots, and consignees and notified their consignees by email on 06/09/2023. Upon receipt of the submitted response form each account will receive over-labels to affix to the affected kits via FedEx overnight, if applicable. distributors were directed to notify their customers and collect the returned product and return it to Medline.
Recall details
- Recall number
- Z-2132-2023
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- May 18, 2023
- Posted by the FDA
- Jul 12, 2023
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 15 units
- Product code
- LRO
- Distribution
- Worldwide distribution - US Nationwide and the country of Panama.
Product description
MEDLINE NON-STERILE PVP SOLUTION, packaged in kits as follows: EAR CDS, Model Number CDS980089O
Code information (models, lots, serial numbers)
UDI/DI (EA) 10195327303969, UDI/DI (CS) 40195327303960; Lot Numbers: 23EBE453, 23CBW365
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. CDS980089OMedline Industries, Inc. · Ear, nose, and throat surgical tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92486).
- Z-2131-2023 · Open, ClassifiedMEDLINE NON-STERILE PVP SOLUTION, STERILE PACKAGING REF DYNDA2232A
- Z-2134-2023 · Open, ClassifiedMEDLINE NON-STERILE PVP SOLUTION, packaged in kits as follows: OISC HAND BEAR, Model Number DYNJ902348L
- Z-2135-2023 · Open, ClassifiedMEDLINE NON-STERILE PVP SOLUTION, packaged in kits as follows: HYSTERO PACK, Model Number DYNJ61448A
- Z-2136-2023 · Open, ClassifiedMEDLINE NON-STERILE PVP SOLUTION, packaged in kits as follows: a) OCULOPLASTY PACK, Model Number DYNJ64572A; b) EYE PACK, Model Number VAL050EEPKB
- Z-2137-2023 · Open, ClassifiedMEDLINE NON-STERILE PVP SOLUTION, packaged in kits as follows: a) TOTAL JOINT, Model Number CDS985555S; b) TOTAL KNEE, Model Number CDS985566Q; c) MAJOR ORTHO-LF, Model Number DYN…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page