Skip to content

HoverMatt, 39" W x 78" L, REF: HM39SPU-B recallZ-2394-2023 · Stryker Sustainability Solutions · initiated Jun 21, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Sustainability Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the patient's weight across, which may increase the effort exerted by staff for a lateral patient transfer, and there may be increased friction and shearing forces when the patient is moved resulting in skin tear injury.

FDA root cause category: Packaging process control

What the company is doing

On 6/21/23, recall notices were distributed to customers who were asked to do the following: 1) Return affected devices. 2) Complete and return the Business Reply form to [email protected] 3) If any of the affected product have been forwarded to additional facilities (including but not limited to distribution warehouses), contact these facilities and communicate this recall to them and instruct them to complete and return the Business Reply Form. Adverse reactions or quality problems can be reported to the Complaint Hotline: +1-888-763-8803 or http://www.stryker.com/productexperience

Recall details

Recall number
Z-2394-2023
Status
Open, Classified
Initiated by the company
Jun 21, 2023
Posted by the FDA
Aug 11, 2023
Recalling firm
Stryker Sustainability Solutions
Quantity
340
Product code
FRZ
Distribution
US: WI, ID, CA, FL, HI, MD, RI, AZ, MA, GA, IL

Product description

HoverMatt, 39" W x 78" L, REF: HM39SPU-B

Code information (models, lots, serial numbers)

UDI-DI: 07613327487718, Lot: 0000185127, Distribution Dates: 11/28/2022-12/08/2022

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates