MAHURKAR recallZ-2371-2023 · Covidien LP · initiated Jun 28, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
FDA root cause category: Process control
What the company is doing
Medtronic issued Urgent Medical Device Recall letter beginning on June 28, 2023. Letter states reason for recall, health risk and action to take: To help you identify if you have affected product, please visit our website www.Medtronic.com/Mahurkar-Triple-Lumen-Catheter-Recall. Here you will find a tool to help you determine if the product you have is affected by this recall. 2. Immediately quarantine and discontinue use of all unused Mahurkar TM Acute Triple Lumen Catheters and Mahurkar TM Acute High Pressure Triple Lumen Catheters referenced in Attachment B - List of affected Lot numbers (see Attachment A for guidance to identify impacted product). Note: This recall does not include Mahurkar Elite Catheters. 3. Complete the online Customer Confirmation Form using the link below within 3 business days of receiving this notice. Replying promptly will confirm your receipt of the notification and prevent you from receiving repeat notices. Link: https://www.novasyte.com/medtronic/fa1333 Note: Please complete the online Customer Confirmation Form even if you DO NOT have affected product. 4. If you have affected product, it must be returned for processing. Upon completion of the online Customer Confirmation Form, Customer Service will contact you with an RGA number to return all unused affected Mahurkar TM Acute Triple Lumen Catheters and Mahurkar TM Acute High Pressure Triple Lumen Catheters from your inventory to Medtronic. 5. This notice should be passed on to all those who need to be aware within your organization or to any organization including but not limited to Nephrologists, Intensivists, physicians, renal nurses, critical care nurses, or other dialysis staff where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Medtronic has partnered with Novasyte, an IQVIA company, to assist in this recall. For assistance regarding online response processing or product return for this recall, please contact Novas
Recall details
- Recall number
- Z-2371-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 28, 2023
- Posted by the FDA
- Aug 11, 2023
- Recalling firm
- Covidien LP
- Quantity
- 1875 units
- Product code
- NIE
- FDA submissions
- K020089, K102605
- Distribution
- US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
Product description
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, IC Tray REF: 8888102003HP
Code information (models, lots, serial numbers)
GTIN: 20884521127996 LOT Numbers: 2204000239 2227800110
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MAHURKAR 8888102003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
Product lines
Other devices under the same FDA submission
23 devices cleared under K020089, K102605 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MAHURKAR 8888101002HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls1 death8 reports
- MAHURKAR 8888101003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888102004HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls3 reports
- MAHURKAR 8888104004HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls0 reports
- MAHURKAR 8888345629Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall22 reports
- MAHURKAR 8888345611Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall11 reports
- MAHURKAR 8888340629HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall8 reports
- MAHURKAR 8888345603Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall7 reports
- MAHURKAR 8888345611HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall6 reports
- MAHURKAR 8888345629HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall6 reports
- MAHURKAR 8888103002HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall3 reports
- MAHURKAR 8888345637Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED1 recall2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92660).
- Z-2367-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP
- Z-2368-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, IC Tray REF: 8888101002HP
- Z-2369-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, IC Tray REF: 8888101003HP
- Z-2370-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, IC Tray REF: 8888101004HP
- Z-2372-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 8888102004HP
- Z-2373-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, PASS Tray REF: 8888103001HP
- Z-2374-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, PASS Tray REF: 8888103002HP
- Z-2375-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, PASS Tray REF: 8888103004HP
- Z-2376-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, PASS Tray REF: 8888104003HP
- Z-2377-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, PASS Tray REF: 8888104004HP
- Z-2378-2023 · Open, ClassifiedMAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629
- Z-2379-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629HP
- Z-2380-2023 · Open, ClassifiedMAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637
- Z-2381-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637HP
- Z-2382-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions REF: 8888345512HP
- Z-2383-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions REF: 8888345520HP
- Z-2384-2023 · Open, ClassifiedMAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit" REF: 8888345603
- Z-2385-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit REF: 8888345603HP
- Z-2386-2023 · Open, ClassifiedMAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit" REF: 8888345611
- Z-2387-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit REF: 8888345611HP
- Z-2388-2023 · Open, ClassifiedMAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit" REF: 8888345629
- Z-2389-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit REF: 8888345629HP
- Z-2390-2023 · Open, ClassifiedMAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637
- Z-2391-2023 · Open, ClassifiedMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637HP
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page