Skip to content

VITROS Immunodiagnostic Products Troponin I ES Reagent Pack recallZ-2357-2023 · Ortho-Clinical Diagnostics · initiated Jun 5, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.

FDA root cause category: Process control

What the company is doing

Customers were sent a recall notification by mail dated 06/05/2023. The notification that customers immediately discontinue use of affected devices and discard any remaining inventory. Any discarded product will be replaced by QuidelOrtho or credit will be provided. Customers are asked to return the provided Confirmation of Receipt Form no later than 06/12/2023. If affected devices were further distributed customers are asked to forward the provided notification to those downstream customers. This notice is to be posted by all VITROS Systems until the expiration of affected devices, which is 06/10/2023. Customers with any questions may contact Quidel Ortho at 1-800-421-3311.

Recall details

Recall number
Z-2357-2023
Status
Open, Classified
Initiated by the company
Jun 5, 2023
Posted by the FDA
Aug 8, 2023
Recalling firm
Ortho-Clinical Diagnostics
Quantity
4,013 units
Product code
MMI
FDA submissions
K062838, K992366
Distribution
Worldwide distribution - US Nationwide and the country of Canada.

Product description

VITROS Immunodiagnostic Products Troponin I ES Reagent Pack

Code information (models, lots, serial numbers)

Product Code: 6802301; UDI-DI: 10758750002504; Lot No.: 4800; Expiration Date: 10-Jun-2023.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 92662).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page