Randox Liquid Urine Control Level 2- intended for in vitro… recallZ-2446-2023 · Randox Laboratories Ltd. · initiated Jun 21, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Randox Laboratories Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.
FDA root cause category: Under Investigation by firm
What the company is doing
Randox Laboratories issued Medical Device Removal letter to the distribution center within the USA and P. Rico dated June 28, 2023. The distributor will then contact the customer directly via email. Letter states reason for recall, health risk and action to take: Action to be taken: " Discontinue use of and discard any of the above immediately. Provide Randox with photographic evidence of the destruction of the kits. " Review your inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. " Review previous results for this controls and ensure patient results were not reported if control results were not within range. " Complete and return the response form 12187-QA to [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation.
Recall details
- Recall number
- Z-2446-2023
- Status
- Open, Classified
- Initiated by the company
- Jun 21, 2023
- Posted by the FDA
- Aug 23, 2023
- Recalling firm
- Randox Laboratories Ltd.
- Quantity
- 39 units
- Product code
- JJY
- FDA submissions
- K110904
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AR, FL, GA, OH, VA including PR and WW distribution.
Product description
Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol, Creatinine, Glucose, hCG pregnancy, Magnesium, Microalbumin, Osmolality, pH, Phosphate Inorganic, Potassium, Protein Total, Sodium, Specific Gravity, Urea and Uric Acid (Urate) on clinical chemistry systems Catalog Number: UC5074
Code information (models, lots, serial numbers)
GTIN: 05055273207569 Batch/Lot Number: 1209UC Expiry Date: 28 Mar 24
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- LIQUID URINE CONTROL - LEVEL 2 (LIQ URN CONTROL 2) UC5074Randox Laboratories Limited · Multi-Analyte Controls, All Kinds (Assayed)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page