Skip to content

SURGICOUNT recallZ-2463-2023 · Stryker Corporation · initiated Jul 10, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for current software to miscount when scanning in multiple sponge-products from the same unique sponge-pack type

FDA root cause category: Under Investigation by firm

What the company is doing

Stryker issued Urgent Medical Device Correction Letter addressed to: IT Director, Materials Manager, Risk Manager, OR Director, OR Coordinator sent July 10, 2023 via 2-day Fed'X delivery. Letter states reason for recall, health risk and action to take: 1. Immediately review this software update notification. Advise OR staff, and all other personnel who may use this product, of the information in the notification. 2. Ensure all Stryker surgical tablets using Surgicount+ software is connected to the local facility Wi-Fi so that the software update can be automatically transmitted. Press INSTALL once the update notification appears on the tablet to complete the update. Contact your local sales representative or technical support at 1-800-253-3210 for any assistance or questions. 3. Sign and return the enclosed Business Reply Form to [email protected]. 4. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. We apologize for any inconvenience this action may cause your facility. Please forward a copy of this letter to any other personnel within your facility that you deem appropriate. Please contact Jordan Anderson at (269) 800-1347 or [email protected] with questions or concerns.

Recall details

Recall number
Z-2463-2023
Status
Open, Classified
Initiated by the company
Jul 10, 2023
Posted by the FDA
Aug 25, 2023
Recalling firm
Stryker Corporation
Quantity
57 units
Product code
PVZ
Distribution
US Nationwide distribution.

Product description

SurgiCount+ Software Application-indicated as an adjunctive technology for augmenting the manual process of counting, displaying and recording the number of RFID-tagged absorbent articles used during surgical procedures Product Code: 0694-002-090;

Code information (models, lots, serial numbers)

GTIN 07613327543094 Software Version: 2.1.8 & 2.20

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page