Skip to content

ProPlan CMF recallZ-2287-2023 · Materialise N.V. · initiated Jun 19, 2023

Completed

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Materialise N.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Wrong versions of guides 3A and 3B (Patient Specific) were shipped prior to the re-design request

FDA root cause category: Process change control

What the company is doing

Synthes sales representative was made aware of the issue on 19 June 2023, by phone and email. Materialise requested the product to be returned on 19 June 2023, Synthes sales representative confirmed that surgery will be performed without guides and will return the case. The guides were returned/received at Materialise on 23 June 2023.

Recall details

Recall number
Z-2287-2023
Status
Completed
Initiated by the company
Jun 19, 2023
Posted by the FDA
Aug 3, 2023
Recalling firm
Materialise N.V.
Quantity
1 unit
Product code
JEY
FDA submissions
K103136
Distribution
US Nationwide distribution in the state of KY.

Product description

ProPlan CMF Patient-Specific Guide, Mandible guides (3A & 3B)-Intended to guide the marking of bone and/or guide surgical instruments in mandibular and maxillofacial surgical procedures. Identification number: SD900.101

Code information (models, lots, serial numbers)

UDI: 05420060351013 Lot Case Number: MU23-JOQ-LID

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under K103136 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page