Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x… recallZ-2488-2023 · Beckman Coulter Inc. · initiated Jul 12, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lipemic interference failed to meet the performance specification listed within the IFU.
FDA root cause category: Other
What the company is doing
On July 18, 2023 Beckman Coulter issued a "Urgent Medical Device Recall" Notification via email and postal mail. Beckman Coulter asked consignees to take the following action: 1. No retrospective re view of past patient results is required however Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. 2. Discontinuance or disposal of this product is not necessary. 3. Per the IFU, avoid highly lipemic samples when using the IgA assay. 4. No update is required to the LIH influences check settings on the analyzer, where in use. 5. Please share this information with your laboratory staff nd retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 6. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: - Electronically - If you received this communication via email, Manually - complete and return the enclosed Response Form. 7. If you have any questions regarding this notice, please contact our Customer Support Center - www.beckmancoulter.com or 1-800-854-3633 8. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Recall details
- Recall number
- Z-2488-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 12, 2023
- Posted by the FDA
- Aug 31, 2023
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 12,850 units
- Product code
- CFN
- FDA submissions
- K073489
- Distribution
- US: AL AR AZ CA CO CT DC DE FL GA IA IL IN KS KY LA MA MD MI MN MO MS NC ND NE NJ NM NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV OUS: Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, C¿te d'Ivoire, Croatia, Cura¿ao, Cyprus, Czechia, Denmark, Ecuador, Egypt, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Korea, Republic of, Lebanon, Lithuania, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Netherlands, New Zealand, Oman, Pakistan, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Province of China, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uruguay, Viet Nam, Yemen.
Product description
Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x 11 mL R2
Code information (models, lots, serial numbers)
All Lots/UDI-(01)15099590011574
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AU OSR61171Beckman Coulter, Inc. · Method, Nephelometric, Immunoglobulins (G, A, M)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.