IMPIX 3D Print recallZ-2452-2023 · Medicrea International · initiated Jul 12, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medicrea International or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for a product packaging non-conformity issue which presents as a pinhole in either the inner or outer pouch.
FDA root cause category: Under Investigation by firm
What the company is doing
Beginning 12 July 2023, In the U.S. Medtronic disseminated an URGENT: MEDICAL PRODUCT RECALL to its consignees via UPS 2-Day delivery. The notice explained the problem, risk, and requested the following: Medtronic requests that you immediately take the following actions: Identify and quarantine any unused impacted product(s). Return all unused and non-expired product(s) in your inventory to Medtronic following the instructions in the enclosed Customer Confirmation Form. Your Medtronic Sales Representative can assist in returning any affected consignment and loaner inventory, if applicable. Complete the Customer Confirmation Form enclosed with this letter (even if you have no product to return), acknowledging that you have received this information. This notice should be distributed to all others in your organization who should be aware, or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records.
Recall details
- Recall number
- Z-2452-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 12, 2023
- Posted by the FDA
- Aug 24, 2023
- Recalling firm
- Medicrea International
- Quantity
- 248 units
- Product code
- MAX
- FDA submissions
- K163595
- Distribution
- US nationwide
Product description
IMPIX 3D, REF numbers: a) B242C09122810, b) B242C10062810, c) B242C10122810, d) B242C11062810, e) B242C11122810, f) B242C12062810, g) B242C12122810, h) B242C12123310, i) B242C13062810, j) B242C13122810, k) B242C14122810, l) B242C15122810, m) B247C07062410, n) B247C07062810, o) B247C07063210, p) B247C08062410, q) B247C08062810, r) B247C08063210, s) B247C09062410, t) B247C09062810, u) B247C09063210, v) B247C10062410, w) B247C10062810, x) B247C10063210, y) B247C11062410, z) B247C11062810, aa) B247C11063210, bb) B247C12062410, cc) B247C12062810, dd) B247C12063210, ee) B247C13062410, ff) B247C13062810, gg) B247C13063210, hh) B247C14062410, ii) B247C14062810, jj) B247C14063210, kk) B247C15062810, ll) B247C15063210; intervertebral fusion device - lumbar
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
a) B242C09122810, GTIN 03613720284444, Lot Numbers: 20B0926, 20B0984; b) B242C10062810, GTIN 03613720270522, Lot Numbers: 18E0499; c) B242C10122810, GTIN 03613720284451, Lot Numbers: 20B0928, 20B0985; d) B242C11062810, GTIN 03613720270539, Lot Numbers: 18D0708, 19F0261; e) B242C11122810, GTIN 03613720284468, Lot Numbers: 20B0930, 20B0986; f) B242C12062810, GTIN 03613720270546, Lot Numbers: 18D0709, 19G0038; g) B242C12122810, GTIN 03613720284475, Lot Numbers: 20B0931, 20B0987; h) B242C12123310, GTIN 03613720284543, Lot Numbers: 20C0374; i) B242C13062810, GTIN 03613720270553, Lot Numbers: 18D0710, 19F0263; j) B242C13122810, GTIN 03613720284482, Lot Numbers: 20E0343; k) B242C14122810, GTIN 03613720284499, Lot Numbers: 20B0932; l) B242C15122810, GTIN 03613720284505, Lot Numbers: 20B0933; m) B247C07062410, GTIN 03613720272762, Lot Numbers: 18D0965, 18I0576, 18L0085; n) B247C07062810, GTIN 03613720272779, Lot Numbers: 18I0577, 18L0084; o) B247C07063210, GTIN 03613720272786, Lot Numbers: 18I0492; p) B247C08062410, GTIN 03613720272533, Lot Numbers: 18D0966, 18I0579, 18L0086; q) B247C08062810, GTIN 03613720272526, Lot Numbers: 18I0554, 18I0580; r) B247C08063210, GTIN 03613720272540, Lot Numbers: 18I0493; s) B247C09062410, GTIN 03613720272557, Lot Numbers: 18D0967, 18I0582, 18L0088, 19F0264; t) B247C09062810, GTIN 03613720272564, Lot Numbers: 18H0611R, 18H0611R/1, 18I0555; u) B247C09063210, GTIN 03613720272571, Lot Numbers: 18I0494; v) B247C10062410, GTIN 03613720272588, Lot Numbers: 18D0968, 18I0584, 18L0089, 20A0451; w) B247C10062810, GTIN 03613720272595, Lot Numbers: 18H0612, 18H0612/1, 18H0612/2; x) B247C10063210, GTIN 03613720272601, Lot Numbers: 18I0495; y) B247C11062410, GTIN 03613720272618, Lot Numbers: 18D0969, 18I0586, 19F0265; z) B247C11062810, GTIN 03613720272625, Lot Numbers: 18H0613, 18H0613/1; aa) B247C11063210, GTIN 03613720272632, Lot Numbers: 18I0496; bb) B247C12062410, GTIN 03613720272649, Lot Numbers: 18D0970; cc) B247C12062810, GTIN 03613720272656, Lot Numbers: 18H0614, 18I0558; dd) B247C12063210, GTIN 03613720272663, Lot Numbers: 18I0560; ee) B247C13062410, GTIN 03613720272670, Lot Numbers: 18I0589; ff) B247C13062810, GTIN 03613720272687, Lot Numbers: 18H0615; gg) B247C13063210, GTIN 03613720272694, Lot Numbers: 18I0561; hh) B247C14062410, GTIN 03613720272700, Lot Numbers: 18L0099; ii) B247C14062810, GTIN 03613720272717, Lot Numbers: 18H0616/1/1, 18H0616/2; jj) B247C14063210, GTIN 03613720272724, Lot Numbers: 18I0562; kk) B247C15062810, GTIN 03613720272748, Lot Numbers: 18I0498; ll) B247C15063210, GTIN 03613720272755, Lot Numbers: 18I0563
Devices named in this recall
38 devices in the FDA device registry whose identifier appears in the recall's code information.
- IMPIX 3D Print B242C12062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall1 report
- IMPIX 3D Print B242C10062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C11062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C13062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C08062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C08062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C08063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C09062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C09062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C09063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C10062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C10062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C10063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C11062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C11062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C11063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C12062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C12062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C12063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C13062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C13062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C13063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C14062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C14062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C14063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C15062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C15063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C07062410Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C07062810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B247C07063210Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C09122810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C10122810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C11122810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C12122810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C13122810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C14122810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C15122810Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
- IMPIX 3D Print B242C12123310Medicrea International · Intervertebral Fusion Device With Bone Graft, Lumbar1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 92738).
- Z-2453-2023 · Open, ClassifiedIMPIX ALIF, REF numbers: a) B15111209, b) B15111212, c) B15111409, d) B15111412; intervertebral fusion device - lumbar
- Z-2455-2023 · Open, ClassifiedIMPIX DLIF, REF numbers: a) B16123509, b) B16123511, c) B16124009, d) B16124013, e) B16124509, f) B16124511, g) B16124513, h) B16164009, i) B16164013, j) B16165009, k)…
- Z-2456-2023 · Open, ClassifiedIMPIX MANTA, REF numbers: a) A20150407, b) A20250407, c) A20350567, d) B20181743, e) B20181753, f) B20181763, g) B20181773, h) B20181943, i) B20181953, j) B20181963, k)…
- Z-2457-2023 · Open, ClassifiedIMPIX TLIF, REF numbers: a) B15130607S, b) B15130609S, c) B15130611S, d) B15130613S, e) B15130615S, f) B15900806, g) B15900906, h) B15901006, i) B15901106, j) B15901206,…
- Z-2458-2023 · Open, ClassifiedPASS LP, REF numbers: a) B02315535Z, b) B02315540Z, c) B02315545Z, d) B02315550Z, e) B02316535Z, f) B02316540Z, g) B02316545Z, h) B02316550Z, i) B02317540Z, j) B02317545…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page