Skip to content

LINK Instrument recallZ-2348-2023 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Jul 3, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue

FDA root cause category: Device Design

What the company is doing

LinkBio Corp initiated Urgent Medical Device Recall Letter to US Distributors via email on July 6, 2023. Letter states reason for recall, health risk and action to take: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. These trials are contained in the MP Monoblock instrument sets. The entire instrument set must be returned as part of this recall. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address, referencing QT 2023-238: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or [email protected] for any questions

Recall details

Recall number
Z-2348-2023
Status
Open, Classified
Initiated by the company
Jul 3, 2023
Posted by the FDA
Aug 4, 2023
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Quantity
24 units
Product code
LZO
FDA submissions
K183141
Distribution
US Nationwide distribution.

Product description

LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 17 - indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-117/00

Code information (models, lots, serial numbers)

UDI-DI: 04026575182114 Lot Numbers: B922124 B932051 C005064 C211011

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

12 devices cleared under K183141 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92758).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page