O-ARM recallZ-2449-2023 · Medtronic Navigation, Inc.-Littleton · initiated Jul 25, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Navigation, Inc.-Littleton or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ground cable installed incorrectly.

FDA root cause category: Under Investigation by firm

What the company is doing

Customers were sent recall notification via UPS 2-day delivery on 7/25/2023. The recall notification informs consignees they may continue to use the device per facility protocol; there is no risk to the patient or user when system panels remain intact. Consignees are to inform all personnel in all care environments in with the recalled device is used about the provided Medical Device Correction Notice and to post copies of the notification in a prominent location to maintain awareness until the issue is resolved by a Medtronic Field Service Engineer. Consignees are to complete and return the provided customer confirmation form via email to [email protected]. If consignees are aware of any incidents related to this recall, they are to contact Technical Support immediately by phone at 1-888-826-5603 or by mail at [email protected]. Medtronic will schedule a service visit to correct this issue but if consignees have any questions regarding this issue in the interim, contact Technical Services for help at 1-800-595-9709 (US) or 1-700-890-3160 (OUS) and reference FA1351.

Recall details

Recall number
Z-2449-2023
Status
Open, Classified
Initiated by the company
Jul 25, 2023
Posted by the FDA
Aug 24, 2023
Recalling firm
Medtronic Navigation, Inc.-Littleton
Quantity
28 units
Product code
OWB
FDA submissions
K151000
Distribution
Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KY, MI, MO, NC, NV, OH, TN, UT, VA, & WA. The countries of Canada, Ireland, Poland, Russian Federation, Switzerland, & Viet Nam.

Product description

The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs. or greater and having an abdominal thickness greater than 16 cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.

Code information (models, lots, serial numbers)

Model No. BI70002000; UDI-DI: 00763000074128, 00763000081041, 00763000355555, 00763000496784, 00763000542801, 00763000616434, 00763000616526, 00763000616564; Serial No. C2786, C2841, C2781, C2624, C2659, C2660, C2665, C2672, C2708, C2723, C2735, C2760, C2774, C2775, C2779, C2798, C2814, C2823, C2873, C2918, C2968, C3290, C3363, C3188, C3322, C3387, C3404, and C3418.