O-ARM recallZ-2449-2023 · Medtronic Navigation, Inc.-Littleton · initiated Jul 25, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Navigation, Inc.-Littleton or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Ground cable installed incorrectly.
FDA root cause category: Under Investigation by firm
What the company is doing
Customers were sent recall notification via UPS 2-day delivery on 7/25/2023. The recall notification informs consignees they may continue to use the device per facility protocol; there is no risk to the patient or user when system panels remain intact. Consignees are to inform all personnel in all care environments in with the recalled device is used about the provided Medical Device Correction Notice and to post copies of the notification in a prominent location to maintain awareness until the issue is resolved by a Medtronic Field Service Engineer. Consignees are to complete and return the provided customer confirmation form via email to [email protected]. If consignees are aware of any incidents related to this recall, they are to contact Technical Support immediately by phone at 1-888-826-5603 or by mail at [email protected]. Medtronic will schedule a service visit to correct this issue but if consignees have any questions regarding this issue in the interim, contact Technical Services for help at 1-800-595-9709 (US) or 1-700-890-3160 (OUS) and reference FA1351.
Recall details
- Recall number
- Z-2449-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 25, 2023
- Posted by the FDA
- Aug 24, 2023
- Recalling firm
- Medtronic Navigation, Inc.-Littleton
- Quantity
- 28 units
- Product code
- OWB
- FDA submissions
- K151000
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KY, MI, MO, NC, NV, OH, TN, UT, VA, & WA. The countries of Canada, Ireland, Poland, Russian Federation, Switzerland, & Viet Nam.
Product description
The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs. or greater and having an abdominal thickness greater than 16 cm and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.
Code information (models, lots, serial numbers)
Model No. BI70002000; UDI-DI: 00763000074128, 00763000081041, 00763000355555, 00763000496784, 00763000542801, 00763000616434, 00763000616526, 00763000616564; Serial No. C2786, C2841, C2781, C2624, C2659, C2660, C2665, C2672, C2708, C2723, C2735, C2760, C2774, C2775, C2779, C2798, C2814, C2823, C2873, C2918, C2968, C3290, C3363, C3188, C3322, C3387, C3404, and C3418.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- O-ARM® BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile2 recalls1 death811 reports
- O-arm BI70002000Medtronic Navigation, Inc. · Interventional fluoroscopic x-ray system2 recalls466 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page