Atellica CH Total Bilirubin_2 -In vitro diagnostic use in… recallZ-2471-2023 · Siemens Healthcare Diagnostics, Inc. · initiated Jul 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica CH Total Bilirubin_2 (TBil_2), AtellicaCH LDL Cholesterol (LDLC), Atellica CH Gamma-Glutamyl Transferase (GGT) and AtellicaCH HDL Cholesterol (HDLC)
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthineers issued Urgent Medical Device Correction ACHC23-01.C.US to US affected customers via FedEx beginning on 07/10/2023; and Urgent Field Safety Notice ACHC23-01.C.OUS,OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 07/10/2023. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Perform the instructions provided in the Additional Information section. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. If your laboratory has multiple Atellica CH 930 Analyzers, separate the assays as follows: " Perform testing of TBil_2 on a separate analyzer from Glucose Oxidase (GluO) " Perform testing of LDLC on a separate analyzer from Lipase (Lip) " Perform testing of GGT on a separate analyzer from Magnesium (Mg) " Perform testing of HDLC on a separate analyzer from Uric Acid (UA) If you choose not to or are unable to separate the assays as indicated above, batch testing of GluO, Lip, Mg and UA may be performed. Note: When batch testing impacted assays, GluO, Lip, Mg and UA, an RPC2 wash mitigation must be initiated after completion of tests with the assay causing carryover, TBil_2, LDLC, GGT and HDLC, respectivel
Recall details
- Recall number
- Z-2471-2023
- Status
- Open, Classified
- Initiated by the company
- Jul 10, 2023
- Posted by the FDA
- Aug 28, 2023
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 50148 units
- Product code
- JFM
- FDA submissions
- K170065
- Distribution
- US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam
Product description
Atellica CH Total Bilirubin_2 -In vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates Test Code: TBil_2 Siemens Material Number (SMN): 11097531
Code information (models, lots, serial numbers)
Unique Device Identification (UDI): 00630414595818 All Lot Numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH TBil_2 11097531Siemens Healthcare Diagnostics Inc. · ENZYMATIC METHOD, BILIRUBIN2 recalls3 reports
Other devices under the same FDA submission
2 devices cleared under K170065 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ADVIA® Chemistry Total Bilirubin_2 TBIL_2 Reagents 10341113Siemens Healthcare Diagnostics Inc. · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM1 recall0 reports
- ADVIA® Chemistry TBIL_2 Total Bilirubin_2 Reagents 10341115Siemens Healthcare Diagnostics Inc. · ENZYMATIC METHOD, BILIRUBIN1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92800).
- Z-2472-2023 · Open, ClassifiedAtellica CH LDL Cholesterol- In vitro diagnostic use in the quantitative determination of LDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium heparin Test Code…
- Z-2473-2023 · Open, ClassifiedAtellica CH Gamma-Glutamyl Transferase-IIn vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma (lithium heparin Test Code…
- Z-2474-2023 · Open, ClassifiedAtellica CH HDL Cholesterol- In vitro diagnostic use in the quantitative determination of HDL cholesterol in human serum and plasma (lithium heparin, EDTA, sodium Heparin) Test C…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page