Citrine Triple Quad Mass Spectrometer, Part Number 5063684 recallZ-0040-2024 · AB Sciex, LLC · initiated Aug 30, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from AB Sciex, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
FDA root cause category: Component design/selection
What the company is doing
An URGENT: RUO and MEDICAL DEVICE FIELD ACTION notification letter dated 7/20/23 was sent to customers. Actions to be taken by the Customer/User: SCIEX will provide a replacement nut for eligible sources and systems. Instructions will be provided to guide you through the replacement of the nut. The replacement parts with installation instructions are estimated to begin shipping in October 2023. SCIEX is currently notifying affected customers of the upcoming countermeasure. Upon availability, SCIEX will send you the replacement nut with instructions on how to install it on the ion source probe. We sincerely apologize for any inconvenience that this may cause you. SCIEX aims to provide you with products of the highest quality. If you have any questions regarding this notice, please contact https://sciex.com/aboutus/ contact-us or https://sciex.com/support/request-support.
Recall details
- Recall number
- Z-0040-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 30, 2023
- Posted by the FDA
- Oct 5, 2023
- Recalling firm
- AB Sciex, LLC
- Quantity
- 14
- Product code
- DOP
- Distribution
- US Nationwide.
Product description
Citrine Triple Quad Mass Spectrometer, Part Number 5063684
Code information (models, lots, serial numbers)
UDI-DI: 00842365100533; Serial Numbers: DE220001803PL, DE230011804, DE230031806, DE230041807, DE230051808, DE230071901, DE230081903, DE230091903, DE230162003, DE230272105, DE230292105, DE240002106, DE240022111, DE240042203
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Citrine Triple Quad™ MS/MS 5050803AB Sciex (Manufacturing) · Mass Spectrometer, Clinical Use1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92810).
- Z-0038-2024 · Open, Classified4500MD Triple Quad Mass Spectrometer, Part Number 5032522
- Z-0039-2024 · Open, ClassifiedCitrine QTRAP Mass Spectrometer, Part Number 5063685
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page