Skip to content

Preat Implant Prosthetics recallZ-0154-2024 · Preat Corp · initiated Jul 18, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Preat Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The healing abutments contain a dimensional condition that may lead to a gap between the healing abutment and the implant allowing for microleakage

FDA root cause category: Process change control

What the company is doing

On August 4, 2023, Preat Corporation issued a "Urgent Medical Device Recall" notification via: USPS. Preat ask consignees to take the following actions: Product within the scope of the recall shall be discarded or returned to Preat Corporation. Instructions for the return will be provided. Complete and email response form to Nichole Fetterman, E-Mail: [email protected] or mail to Preat Corporation, Attn. Nichole Fetterman, 2625 Skyway Drive, Ste. B, Santa Maria, CA 93455

Recall details

Recall number
Z-0154-2024
Status
Open, Classified
Initiated by the company
Jul 18, 2023
Posted by the FDA
Oct 24, 2023
Recalling firm
Preat Corp
Quantity
4 units
Product code
NHA
FDA submissions
K220823
Distribution
US: FL, CA, OR, TX, OH, CA, SC, OR, NJ, SC, OUS: None

Product description

Preat REF 9007124, Neodent GM x 4mm Healing Abutment (Wide), Rx Only, Non-Sterile, Single Use Only. Dental

Code information (models, lots, serial numbers)

UDI-DI: 00842092161395/Lot #OF018355

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K220823 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92867).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page