FilmArray Gastrointestinal (GI) Panel recallZ-2588-2023 · BioFire Diagnostics, LLC · initiated Apr 25, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE FILMARRAY Gastrointestinal (GI) Panel

FDA root cause category: Software design

What the company is doing

On 4/25/23 correction notices were mailed to customers who were told to update the GI pouch module software. Instructions for installation can be downloaded here: https://www.biofiredx.com/e-labeling/ITIFA20GI21. This technical note and download cannot be opened via Firefox browser; please use a different browser. Once the pouch module software has been updated, complete and return the webform located at: https://www.biofiredx.com/gisoftware-5747 or return the acknowledgement form via email to [email protected] Questions or concerns can be directed to Technical Support at [email protected] or via telephone by dialing 1.800.735.6544 and selecting option 5.

Recall details

Recall number
Z-2588-2023
Status
Open, Classified
Initiated by the company
Apr 25, 2023
Posted by the FDA
Sep 30, 2024
Recalling firm
BioFire Diagnostics, LLC
Quantity
61,069
Product code
PCH
FDA submissions
K140407, K160459, K143005
Distribution
US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LV, MO, MX, MY, NL, NO, NP, NZ, PG, PH, PL, PT, QA, RS, SA, SE, SG, SI, TH, TN, TR, TW, VN, ZA, ZW

Product description

FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software

Code information (models, lots, serial numbers)

Model/UDI-DI: RFIT-ASY-0104/00815381020116, RFIT-ASY-0116/00815381020109 using software versions prior to FLM2-SFW-0040 v2.0.3.1/BFS0001-5539 v2.0.200.10

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page