Skip to content

Aquilex Fluid Control System recallZ-2629-2023 · WOM World Of Medicine AG · initiated Jul 17, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from WOM World Of Medicine AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The display of inflow volume can reach its limit of 30,000 ml during long procdeure and the inflow volume display will freeze at the maximum value while the deficit will start counting backwards until 0 ml is reached and may result in fluid overload

FDA root cause category: Software design

What the company is doing

W.O.M. World of Medicine GmbH (WOM), notified the US/Canada distributor Hologic via E-mail Urgent Field Safety Notice (FSN) - MEDICAL DEVICE RECALL Aquilex Fluid Control System Pump - REF AQL-100PBS and AQL-100P WOM reference no.: 2023-0001 on 7/17/23. Hologic notified sales representatives and the end users (hospitals and medical facilities) in the US and Canada on 8/21/2023 via Telephone calls and letter. Letter states reason for recall, health risk and action to take: W.O.M. WORLD OF MEDICINE GmbH has developed a software version, which, among others, starts emitting visual and acoustic signals when an inflow volume of 28,000 ml is reached. Additionally, the system will pause when an inflow volume of 30,000 ml is reached and is designed to prepare the user for a manual fluid deficit determination. Action To Be Taken by the User: Ensure that the affected devices receive their preventive maintenance on time, every two years, as prescribed by the manufacturer. A new software version and an updated version of the IFU are implemented during the next due preventive maintenance of the pump of the Aquilex Fluid Control Systems AQL-100PBS as well as AQL-100P. Also, please complete Customer Acknowledgement Form (Annex 2). This notice should be provided to all personnel who need to be made aware of this issue within your organization.

Recall details

Recall number
Z-2629-2023
Status
Open, Classified
Initiated by the company
Jul 17, 2023
Posted by the FDA
Jan 5, 2024
Recalling firm
WOM World Of Medicine AG
Quantity
1287 units
Product code
HIG
FDA submissions
K172040
Distribution
Nationwide Foreign: Canada, Australia, New Zealand, Belgium, France, Germany, Ireland, Netherlands, Spain, Sweden, Norway, Switzerland, UK, Kuwait, Qatar, Turkey, UAE

Product description

Aquilex Fluid Control System-Intended to provide fluid distension of the uterus during diagnostic and operative hysteroscopies and to monitor the volume differential between the irrigation fluid flowing into and out of the uterus. Models AQL-100PBS and AQL-100P

Code information (models, lots, serial numbers)

Show the full code information (15k characters)

UDI-DI: 04056702006914 Device Models AQL-100PBS and AQL-100P Serial Numbers: US S1802CE0538 1802CE0539 1802CE0540 1802CE0542 1802CE0543 1802CE0544 1802CE0545 1802CE0546 1802CE0547 1802CE0616 1802CE0617 1802CE0618 1802CE0619 1802CE0620 1802CE0621 1802CE0622 1802CE0624 1802CE0625 1802CE0648 1802CE0649 1802CE0651 1802CE0652 1802CE0653 1802CE0654 1802CE0655 1802CE0656 1802CE0657 1803CE0121 1803CE0122 1803CE0123 1803CE0124 1803CE0125 1803CE0126 1803CE0127 1803CE0128 1803CE0129 1803CE0130 1803CE0199 1803CE0201 1803CE0204 1803CE0205 1803CE0295 1803CE0296 1803CE0298 1803CE0299 1803CE0300 1803CE0301 1803CE0302 1803CE0663 1803CE0909 1803CE0910 1803CE0911 1803CE0912 1803CE0913 1803CE0914 1803CE0915 1803CE0916 1803CE0917 1803CE0918 1803CE0984 1803CE0985 1803CE0986 1803CE0987 1803CE0988 1803CE0989 1803CE0990 1803CE0991 1803CE0992 1803CE1014 1803CE1015 1803CE1016 1803CE1017 1803CE1018 1803CE1019 1803CE1020 1803CE1021 1803CE1022 1803CE1023 1803CE1088 1803CE1089 1803CE1090 1803CE1091 1803CE1092 1803CE1093 1803CE1094 1803CE1095 1803CE1096 1803CE1097 1803CE1129 1803CE1131 1803CE1132 1803CE1134 1803CE1135 1803CE1136 1803CE1137 1803CE1178 1803CE1179 1803CE1180 1803CE1181 1803CE1183 1803CE1185 1803CE1186 1803CE1187 1804CE0133 1804CE0134 1804CE0135 1804CE0137 1804CE0138 1804CE0139 1804CE0140 1804CE0142 1804CE0242 1804CE0243 1804CE0244 1804CE0245 1804CE0246 1804CE0247 1804CE0248 1804CE0249 1804CE0312 1804CE0313 1804CE0314 1804CE0316 1804CE0317 1804CE0318 1804CE0319 1804CE0320 1804CE0321 1804CE0402 1804CE0403 1804CE0404 1804CE0405 1804CE0406 1804CE0407 1804CE0408 1804CE0409 1804CE0410 1804CE0446 1804CE0447 1804CE0448 1804CE0449 1804CE0450 1804CE0452 1804CE0453 1804CE0454 1804CE0455 1804CE0541 1804CE0542 1804CE0543 1804CE0544 1804CE0545 1804CE0546 1804CE0547 1804CE0548 1804CE0549 1804CE0550 1804CE0622 1804CE0623 1804CE0624 1804CE0625 1804CE0626 1804CE0749 1804CE0752 1804CE0755 1804CE0801 1804CE0802 1804CE0803 1804CE0804 1804CE0805 1804CE0806 1804CE0807 1804CE0808 1804CE0809 1804CE0810 1804CE0891 1804CE0892 1804CE0893 1804CE0894 1804CE0895 1804CE0896 1804CE0898 1804CE0899 1804CE0900 1804CE0941 1804CE0943 1804CE0944 1804CE0945 1804CE0946 1804CE0947 1804CE0948 1804CE0949 1804CE0950 1805CE0041 1805CE0042 1805CE0043 1805CE0045 1805CE0046 1805CE0047 1805CE0048 1805CE0049 1805CE0164 1805CE0169 1805CE0170 1805CE0171 1805CE0172 1805CE0178 1805CE0179 1805CE0180 1805CE0181 1805CE0182 1805CE0261 1805CE0262 1805CE0263 1805CE0264 1805CE0266 1805CE0267 1805CE0268 1805CE0422 1805CE0423 1805CE0424 1805CE0425 1805CE0426 1805CE0427 1805CE0428 1805CE0429 1805CE0430 1805CE0500 1805CE0506 1805CE0507 1805CE0509 1805CE0510 1805CE0511 1805CE0512 1805CE0513 1805CE0514 1805CE0515 1805CE0550 1805CE0551 1805CE0552 1805CE0553 1805CE0554 1805CE0555 1805CE0557 1805CE0752 1805CE0753 1805CE0754 1805CE0755 1805CE0756 1805CE0757 1805CE0758 1805CE0759 1805CE0760 1805CE0761 1805CE0936 1805CE0938 1805CE0939 1805CE0940 1805CE0941 1805CE0942 1805CE0943 1805CE0944 1805CE0945 1806CE0185 1806CE0186 1806CE0187 1806CE0188 1806CE0189 1806CE0190 1806CE0191 1806CE0192 1806CE0193 1806CE0194 1806CE0359 1806CE0360 1806CE0361 1806CE0362 1806CE0363 1806CE0364 1806CE0365 1806CE0366 1806CE0367 1806CE0368 1806CE0419 1806CE0421 1806CE0422 1806CE0423 1806CE0424 1806CE0425 1806CE0426 1806CE0427 1806CE0428 1806CE0609 1806CE0611 1806CE0612 1806CE0613 1806CE0614 1806CE0615 1806CE0616 1806CE0617 1806CE0618 1807CE0293 1807CE0294 1807CE0295 1807CE0296 1807CE0297 1807CE0298 1807CE0300 1807CE0301 1807CE0302 1807CE0353 1807CE0355 1807CE0357 1807CE0361 1807CE0395 1807CE0396 1807CE0397 1807CE0398 1807CE0399 1807CE0400 1807CE0401 1807CE0402 1807CE0403 1807CE0404 1808CE1124 1808CE1125 1808CE1126 1808CE1128 1808CE1129 1808CE1221 1808CE1222 1808CE1223 1808CE1224 1808CE1308 1808CE1309 1808CE1310 1808CE1312 1808CE1313 1808CE1314 1808CE1315 1808CE1317 1809CE0241 1809CE0242 1809CE0244 1809CE0245 1809CE0246 1809CE0247 1809CE0248 1809CE0281 1809CE0282 1809CE0283 1809CE0284 1809CE0285 1809CE0286 1809CE0287 1809CE0288 1809CE0332 1809CE0333 1809CE0334 1809CE0335 1809CE0336 1809CE0337 1809CE0338 1809CE0339 1809CE0340 1809CE0341 1809CE0601 1809CE0602 1809CE0603 1809CE0604 1809CE0605 1809CE0606 1809CE0608 1809CE0609 1809CE0610 1809CE0673 1809CE0674 1809CE0675 1809CE0677 1809CE0678 1809CE0679 1809CE0680 1809CE0681 1809CE0682 1809CE0703 1809CE0704 1809CE0705 1809CE0706 1809CE0707 1809CE0708 1809CE0709 1809CE0710 1809CE0711 1809CE0712 1809CE0888 1809CE0889 1809CE0890 1809CE0891 1809CE0892 1809CE0893 1809CE0894 1809CE0895 1809CE0896 1809CE0897 1809CE0995 1809CE0996 1809CE0997 1809CE0998 1809CE0999 1809CE1000 1809CE1001 1809CE1002 1809CE1003 1809CE1049 1809CE1050 1809CE1051 1809CE1052 1809CE1054 1809CE1055 1809CE1056 1809CE1057 1810CE0315 1811CE0770 1811CE0771 1811CE0772 1811CE0773 1811CE0774 1811CE0776 1811CE0777 1811CE0778 1811CE0779 1811CE0780 1811CE0781 1811CE0782 1811CE0843 1811CE0844 1811CE0845 1811CE0846 1811CE0847 1811CE0848 1811CE0849 1811CE0851 1811CE0852 1811CE0901 1811CE0903 1811CE0906 1811CE0907 1811CE0909 1902CE0588 1902CE0589 1902CE0590 1902CE0591 1902CE0594 1902CE0628 1902CE0629 1902CE0630 1902CE0632 1902CE0635 1902CE0636 1902CE0692 1902CE0693 1902CE0694 1902CE0695 1902CE0696 1902CE0697 1902CE0698 1902CE0699 1902CE0700 1902CE0701 1902CE0748 1902CE0749 1902CE0752 1902CE0753 1902CE0755 1902CE0757 1903CE0536 1903CE0538 1903CE0540 1903CE0542 1903CE0555 1903CE0556 1903CE0557 1903CE0560 1903CE0587 1903CE0588 1903CE0592 1903CE0593 1903CE0594 1903CE0595 1903CE0596 1903CE0782 1903CE0784 1903CE0785 1903CE0787 1903CE0788 1903CE0789 1904CE0075 1904CE0076 1904CE0078 1904CE0084 1904CE0108 1904CE0110 1904CE0112 1904CE0114 1904CE0115 1904CE0116 1904CE0751 1904CE0754 1904CE0755 1904CE0758 1904CE0804 1905CE0217 1905CE0218 1905CE0219 1905CE0220 1905CE0221 1905CE0222 1905CE0223 1905CE0224 1905CE0225 1905CE0226 1905CE0419 1905CE0420 1905CE0421 1905CE0422 1905CE0423 1905CE0424 1905CE0425 1905CE0426 1905CE0427 1905CE0428 1905CE1031 1905CE1032 1905CE1033 1905CE1034 1905CE1035 1905CE1036 1905CE1037 1905CE1038 1905CE1039 1905CE1040 1906CE0072 1906CE0073 1906CE0074 1906CE0075 1906CE0076 1906CE0077 1906CE0078 1906CE0079 1906CE0080 1906CE0081 1906CE0213 1906CE0214 1906CE0215 1906CE0216 1906CE0217 1906CE0218 1906CE0219 1906CE0220 1906CE0221 1906CE0222 1906CE0288 1906CE0289 1906CE0290 1906CE0291 1906CE0292 1906CE0293 1906CE0294 1906CE0295 1906CE0296 1906CE0297 1906CE0712 1906CE0714 1906CE0715 1906CE0716 1906CE0719 1906CE0720 1906CE0754 1906CE0756 1906CE0758 1906CE0762 1906CE0946 1906CE0948 1906CE0949 1906CE0950 1906CE0951 1906CE0952 1906CE0953 1906CE0954 1906CE0955 1907CE1050 1907CE1051 1907CE1052 1907CE1053 1907CE1054 1907CE1055 1907CE1056 1907CE1057 1907CE1058 1907CE1059 1908CE0358 1908CE0359 1908CE0360 1908CE0362 1908CE0363 1908CE0364 1908CE0365 1908CE0366 1908CE0367 1908CE0615 1908CE0616 1908CE0617 1908CE0618 1908CE0619 1908CE0620 1908CE0621 1908CE0622 1908CE0623 1908CE0624 1908CE0787 1908CE0788 1908CE0790 1908CE0792 1908CE0794 1908CE0825 1908CE0827 1908CE0828 1908CE0829 1908CE0831 1908CE0832 1908CE0833 1910CE0394 1910CE0395 1910CE0396 1910CE0397 1910CE0399 1910CE0556 1910CE0557 1910CE0558 1910CE0559 1910CE0560 1910CE0561 1910CE0562 1910CE0563 1910CE0564 1910CE0565 1911CE0094 1911CE0095 1911CE0096 1911CE0097 1911CE0098 1911CE1026 1911CE1028 1911CE1029 1911CE1102 1911CE1103 1911CE1104 1911CE1105 1911CE1106 1911CE1107 1911CE1108 1911CE1109 1911CE1110 1911CE1111 1911CE1149 1911CE1150 1911CE1151 1911CE1152 1911CE1155 1911CE1158 1911CE1314 1911CE1316 1911CE1318 1912CE0085 1912CE0086 1912CE0087 1912CE0090 1912CE0092 1912CE0093 1912CE0094 1912CE0353 1912CE0354 1912CE0357 1912CE0360 2002CE0987 2002CE0988 2002CE0989 2002CE0990 2002CE0991 2002CE0992 2002CE1060 2006CE0057 2006CE0058 2006CE0059 2006CE0060 2006CE0062 2006CE0276 2006CE0277 2006CE0278 2006CE0279 2006CE0477 2006CE0478 2006CE0479 2006CE0480 2006CE0481 2006CE0482 2006CE0483 2006CE0484 2006CE0485 2006CE0486 2006CE0529 2006CE0530 2006CE0531 2006CE0532 2006CE0533 2006CE0534 2006CE0535 2006CE0536 2006CE0537 2006CE0538 2006CE0877 2006CE0878 2006CE0879 2006CE0881 2006CE0882 2006CE0883 2006CE0884 2007CE0497 2007CE0498 2007CE0499 2007CE0500 2007CE0501 2007CE0502 2007CE0503 2007CE0504 2007CE0505 2007CE0506 2008CE0106 2008CE0107 2008CE0110 2008CE0111 2008CE0112 2008CE0113 2008CE0115 2008CE0206 2008CE0208 2008CE0209 2008CE0211 2008CE0212 2008CE0213 2008CE0214 2008CE0215 2008CE0556 2008CE0557 2008CE0558 2008CE0559 2008CE0560 2008CE0561 2008CE0562 2008CE0563 2008CE0564 2008CE0565 2010CE0985 2010CE0986 2010CE0987 2010CE0988 2010CE0989 2010CE0990 2010CE0992 2010CE0993 2010CE0994 2012CE0711 2012CE0712 2012CE0713 2012CE0714 2012CE0718 2012CE0720 2012CE0721 2102CE0322 2102CE0323 2102CE0324 2102CE0325 2102CE0327 2102CE0328 2102CE0329 2102CE0330 2102CE0331 2102CE0362 2102CE0363 2102CE0364 2102CE0365 2102CE0366 2102CE0367 2102CE0368 2102CE0369 2102CE0370 2102CE0371 2103CE0350 2103CE0351 2103CE0352 2103CE0353 2103CE0354 2103CE0356 2103CE0357 2103CE0358 2103CE0359 2103CE0435 2103CE0436 2103CE0437 2103CE0438 2103CE0439 2103CE0440 2103CE0441 2103CE0442 2103CE0443 2103CE0444 2103CE0800 2103CE0801 2103CE0802 2103CE0803 2103CE0804 2103CE0805 2103CE0807 2104CE0705 2104CE0708 2104CE0709 2104CE0711 2104CE0712 2104CE0713 2104CE0714 2104CE0832 2104CE0833 2104CE0834 2104CE0835 2104CE0836 2104CE0837 2104CE0838 2104CE0839 2104CE0840 2104CE0841 2105CE0403 2105CE0404 2105CE0405 2105CE0406 2105CE0408 2105CE0409 2105CE0410 2105CE0411 2105CE0597 2105CE0598 2105CE0599 2105CE0600 2105CE0601 2105CE0603 2105CE0604 2105CE0605 2105CE0744 2105CE0745 2105CE0746 2105CE0747 2105CE0748 2105CE0749 2105CE0750 2105CE0752 2105CE0753 2106CE0094 2106CE0096 2106CE0097 2106CE0098 2106CE0099 2106CE0100 2106CE0101 2106CE0102 2106CE0103 2107CE0256 2107CE0257 2107CE0258 2107CE0259 2107CE0260 2107CE0261 2107CE0262 2107CE0263 2107CE0264 2107CE0265 2107CE0511 2107CE0513 2107CE0514 2107CE0515 2108CE0551 2108CE0553 2108CE0554 2108CE0555 2108CE0556 2108CE0557 2108CE0559 2108CE0560 2108CE0727 2108CE0728 2108CE0729 2108CE0730 2108CE0731 2108CE0732 2108CE0733 2108CE0734 2108CE0735 2109CE0215 2109CE0216 2109CE0217 2109CE0218 2109CE0219 2109CE0220 2109CE0221 2109CE0222 2109CE0223 2109CE0224 2109CE1026 2109CE1027 2109CE1029 2109CE1032 2109CE1034 2109CE1035 2111CE0547 2111CE0548 2111CE0549 2111CE0550 2111CE0551 2111CE0552 2111CE0553 2111CE0554 2111CE0555 2111CE0556 2111CE0870 2111CE0871 2111CE0872 2111CE0873 2111CE0874 2111CE0875 2111CE0876 2111CE0877 2111CE0921 2111CE0922 2111CE0923 2111CE0925 2111CE0926 2111CE0928 2112CE0101 2112CE0102 2112CE0103 2112CE0104 2112CE0105 2112CE0106 2112CE0107 2112CE0108 2112CE0109 2112CE0110 2112CE0764 2112CE0765 2112CE0769 2112CE0821 2112CE0822 2112CE0823 2112CE0824 2112CE0826 2112CE0849 2112CE0850 2112CE0851 2112CE0852 2112CE0853 2112CE0854 2112CE0855 2112CE0856 2112CE0857 2112CE0858 2112CE0859 2201CE0360 2201CE0361 2204CE0474 2204CE0475 2204CE0476 2204CE0477 2204CE0478 2204CE0479 2204CE0480 2204CE0481 2204CE0482 2204CE0483 2205CE0053 2205CE0054 2205CE0056 2205CE0057 2205CE0058 2205CE0059 2205CE0060 2205CE0061 2205CE0062 2206CE0457 2206CE0458 2206CE0459 2206CE0460 2206CE0461 2206CE0462 2206CE0463 2206CE0464 2206CE0466 2209CE0766 2209CE0767 2209CE0768 2209CE0769 2209CE0770 2209CE0771 2209CE0772 2209CE0773 2209CE0774 2209CE0806 2209CE0807 2209CE0808 2209CE0809 2209CE0810 2209CE0891 2209CE0895 2209CE0896 2209CE0897 1803CE0293 1805CE0166 2103CE0806 2104CE0710 2105CE0407 2105CE0602 2105CE0751 2106CE0095 2107CE0508 2107CE0510 2107CE0516 2107CE0517 2108CE0552 2108CE0558 2108CE0736 2109CE1028 2109CE1030 2109CE1033 2201CE0359 2209CE0813 2209CE0814 2209CE0892 2107CE0510 2109CE1030 2111CE0879 2111CE0924 2206CE0465 2209CE0776 2209CE0777 2209CE0778 2209CE0780 2209CE0781 2209CE0782 2209CE0783 2209CE0784 2209CE0785 2209CE0811 2209CE0812 2209CE0815 2209CE0828 2209CE0829 2209CE0830 2209CE0831 2209CE0832 2209CE0878 2209CE0879 2209CE0880 2209CE0881 2209CE0882 2209CE0883 2209CE0885 2209CE0886 2209CE0887 2209CE0888 2209CE0889 2209CE0890 2209CE0893 2209CE0894 2209CE0916 2209CE0917 2209CE0918 2209CE0919 2209CE0920 2209CE0921 2209CE0922 2209CE0923 2209CE0924 2209CE0925 2209CE0956 2209CE0957 2209CE0958 2209CE0959 2209CE0960 2209CE0961 2209CE0962 2209CE0963 2209CE0964 2209CE0965 2209CE0986 2209CE0987 2209CE0988 2209CE0989 2209CE0990 2209CE0991 2209CE0992 2209CE0993 2209CE0994 2209CE0995 2210CE0089 2210CE0090 2210CE0091 2210CE0092 2210CE0093 2210CE0094 2210CE0095 2210CE0096 2210CE0097 2210CE0098 2210CE0143 2210CE0144 2210CE0145 2210CE0146 2210CE0147 2210CE0148 2210CE0149 2210CE0150 2210CE0151 2210CE0152 2210CE0201 2210CE0202 2210CE0203 2210CE0204 2210CE0205 2210CE0206 2210CE0207 2210CE0208 2210CE0209 2210CE0210 2210CE0811 2210CE0812 2210CE0813 2210CE0814 2210CE0815 2210CE0816 2210CE0817 2210CE0818 2210CE0819 2210CE0820 2211CE0227 2211CE0228 2211CE0229 2211CE0230 2211CE0231 2211CE0232 2211CE0233 2211CE0234 2211CE0235 2211CE0236 2211CE0272 2211CE0273 2211CE0274 2211CE0275 2211CE0276 2211CE0277 2211CE0278 2211CE0279 2211CE0280 2211CE0281 2212CE0356 2212CE0357 2212CE0358 2212CE0359 2212CE0360 2212CE0361 2212CE0362 2212CE0363 2212CE0364 2212CE0365 2212CE0484 2212CE0485 2212CE0486 2212CE0487 2212CE0488 2212CE0489 2212CE0490 2212CE0491 2212CE0492 2212CE0493 2212CE0531 2212CE0532 2212CE0533 2212CE0534 2212CE0535 2212CE0536 2212CE0537 2212CE0538 2212CE0539 2212CE0540 2301CE0613 2301CE0614 2301CE0615 2301CE0616 2301CE0617 2301CE0618 2301CE0619 2301CE0620 2301CE0621 2301CE0622 2301CE0635 2301CE0636 2301CE0637 2301CE0638 2301CE0639 2301CE0640 2301CE0641 2301CE0642 2301CE0643 2301CE0644 2301CE0665 2301CE0666 2301CE0667 2301CE0668 2301CE0669 2301CE0670 2301CE0671 2301CE0672 2301CE0673 2301CE0674 2302CE0386 2302CE0387 2302CE0388 2302CE0389 2302CE0390 2302CE0391 2302CE0392 2302CE0393 2302CE0394 2302CE0395 2302CE0437 2302CE0438 2302CE0439 2302CE0440 2302CE0441 2302CE0442 2302CE0443 2302CE0444 2302CE0445 2302CE0446 2303CE0455 2303CE0456 2303CE0457 2303CE0458 2303CE0459 2303CE0460 2303CE0461 2303CE0462 2303CE0463 2303CE0464 2303CE0497 2303CE0498 2303CE0499 2303CE0500 2303CE0501 2303CE0502 2303CE0503 2303CE0504 2303CE0505 2303CE0506 2303CE0601 2303CE0602 2303CE0603 2303CE0604 2303CE0605 2303CE0606 2303CE0607 2303CE0608 2303CE0609 2303CE0610 2303CE0651 2303CE0652 2303CE0653 2303CE0654 2303CE0655 2303CE0656 2303CE0657 2303CE0658 2303CE0659 2303CE0660 2303CE0661 Serial Numbers: Canada 1802CE0650 1803CE0302 1803CE0662 1803CE1128 1803CE1184 1804CE0136 1804CE0141 1804CE0618 1804CE0627 1804CE0753 1804CE0754 1804CE0756 1805CE0044 1805CE0175 1805CE0260 1805CE0265 1805CE0558 1805CE0559 1807CE0299 1807CE0354 1807CE0356 1807CE0358 1808CE1123 1808CE1127 1808CE1130 1808CE1229 1809CE0249 1809CE1058 1811CE0775 1811CE0900 1811CE0902 1811CE0904 1811CE0906 1902CE0595 1902CE0597 1902CE0634 1902CE0754 1902CE0756 1903CE0535 1903CE0563 1903CE0790 1904CE0752 1904CE0753 1904CE0755 1904CE0756 1904CE0757 1904CE0802 1904CE0805 1904CE0806 1904CE0807 1904CE0809 1904CE0810 1906CE0361 1906CE0362 1906CE0365 1906CE0366 1906CE0713 1906CE0717 1906CE0718 1906CE0755 1906CE0757 1906CE0759 1906CE0761 1906CE0763 1906CE0947 1908CE0824 1908CE0830 1911CE1153 1911CE1154 1911CE1156 1912CE0358 1912CE0359 1912CE0361 2002CE1061 2002CE1062 2002CE1063 2002CE1064 2002CE1065 2002CE1066 2002CE1067 2006CE0063 2006CE0876 2006CE0885 2008CE0109 2010CE0991 2102CE0326 2103CE0809 2104CE0706

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K172040 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page