Skip to content

BD Cathena Safety IV Catheter recallZ-0025-2024 · Becton Dickinson Infusion Therapy Systems Inc. · initiated Aug 29, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson Infusion Therapy Systems Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.

FDA root cause category: Process change control

What the company is doing

On August 29, 2023, Becton Dickinson issued an "Urgent Medical Device Recall" notification via. FedEx. Becton Dickinson asked consignees to take the following actions: 1. Immediately review your inventory for Catalog Number 386862, Lot number 2238464. Destroy all unused product subject to the recall following your institution's process for destruction. 2. For indwelling catheters, no further action is needed. 3. If a defective product was used on a patient and the used needle was disposed of without the clinician suffering a needlestick injury, no further action is necessary to further protect the patient or the clinician. 4. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch, Phone: 1-800-FDA 1088 (1-800-332-1088), Mail: MedWatch, HF-2, FDA, 5600 Fisher.s Lane, Rockville, MD 20852-978

Recall details

Recall number
Z-0025-2024
Status
Open, Classified
Initiated by the company
Aug 29, 2023
Posted by the FDA
Oct 3, 2023
Recalling firm
Becton Dickinson Infusion Therapy Systems Inc.
Quantity
98,000 units
Product code
FOZ
FDA submissions
K220584
Distribution
US Nationwide distribution in the states of CA, FL, GA, MA, MN, MO, MS, NJ, OH, TX.

Product description

REF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO

Code information (models, lots, serial numbers)

Lot # 2238464/ UDI-DI: (01)00382903868629

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates