2008T BlueStar Hemodialysis Machine recallZ-0033-2024 · Fresenius Medical Care Holdings, Inc. · initiated Sep 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential PCBA leaching from tubing of hemodialysis machines.
FDA root cause category: Device Design
What the company is doing
Consignees were sent an URGENT MEDICAL DEVICE CORRECTION NOTICE via USPS, dated 09/06/2023. The notice informed consignees that there is a potential for NDL-PCBAs could leach from silicone tubing manufactured with a peroxide-based catalyst. The firm also states that it has been demonstrated that dialysate run through 2008 Series machines used during typical maintenance HD treatments for at least 36 days, cumulatively, showed levels of NDL-PCBAs that were below toxicologically relevant thresholds. Consignees are to contact FMCRTG Technical Service at 1-800-227-2572 to schedule an update to devices to receive the correct tubing or if they have any further questions. Consignees with clinical questions are to contact Global Medical Information and Education Office at 1-855-2309. Consignees are to return the provided response form via fax or mail.
Recall details
- Recall number
- Z-0033-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 6, 2023
- Posted by the FDA
- Oct 6, 2023
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 733 units
- Product code
- KDI
- FDA submissions
- K150708, K222952, K173972
- Distribution
- Domestic: Nationwide Distribution.
Product description
2008T HD SYS. CDX W/BIBAG BLUESTAR
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Model No: 191126; UDI-DI: 00840861102099; Serial No. 1T0S261032, 1T0S261004, 1T0S261029, 1T0S260995, 1T0S260937, 1T0S260739, 1T0S260687, 1T0S260573, 2T0S263945, 2T0S263947, 2T0S263959, 2T0S263924, 2T0S263962, 2T0S263956, 2T0S263951, 2T0S263950, 2T0S263707, 2T0S263692, 2T0S263690, 2T0S263682, 2T0S263700, 2T0S264084, 2T0S263944, 2T0S263954, 2T0S263957, 2T0S264087, 2T0S264144, 2T0S264132, 2T0S264073, 2T0S264139, 2T0S264061, 2T0S263979, 2T0S263983, 2T0S263973, 2T0S263977, 2T0S263989, 2T0S263938, 2T0S263960, 2T0S263943, 2T0S263968, 2T0S263941, 2T0S263940, 2T0S263939, 2T0S263934, 2T0S263933, 2T0S263932, 2T0S263931, 2T0S263930, 2T0S263927, 2T0S263926, 2T0S263925, 2T0S263923, 2T0S263922, 2T0S263921, 2T0S263920, 2T0S263919, 2T0S263918, 2T0S263916, 2T0S263912, 2T0S263911, 2T0S263910, 2T0S263792, 2T0S263791, 2T0S263784, 2T0S263780, 2T0S263778, 2T0S263776, 2T0S263775, 2T0S263771, 2T0S263769, 2T0S263767, 2T0S263754, 2T0S263752, 2T0S263721, 2T0S263715, 2T0S263711, 2T0S263703, 2T0S263948, 2T0S263718, 2T0S263935, 2T0S263946, 2T0S263953, 2T0S263785, 2T0S263710, 2T0S263615, 2T0S264240, 2T0S264210, 2T0S264246, 2T0S264220, 2T0S264243, 2T0S264250, 2T0S264211, 2T0S264221, 2T0S264247, 2T0S264255, 2T0S264245, 2T0S264248, 2T0S264241, 2T0S264249, 2T0S264264, 2T0S264258, 2T0S264232, 2T0S264234, 2T0S264235, 2T0S264259, 2T0S264265, 2T0S264263, 2T0S264260, 2T0S264227, 2T0S264226, 2T0S264229, 2T0S264224, 2T0S264230, 2T0S264231, 2T0S264236, 2T0S264256, 2T0S264239, 1T0S260442, 1T0S260555, 1T0S260595, 1T0S260603, 1T0S260606, 1T0S260607, 1T0S260609, 1T0S260610, 1T0S260611, 1T0S260612, 1T0S260613, 1T0S260614, 1T0S260615, 1T0S260619, 1T0S260622, 1T0S260627, 1T0S260635, 1T0S260639, 1T0S260641, 1T0S260650, 1T0S260661, 1T0S260678, 1T0S260724, 1T0S260760, 1T0S260768, 1T0S260771, 1T0S260774, 1T0S260783, 2T0S262650, 2T0S262689, 2T0S262694, 2T0S262698, 2T0S262712, 2T0S263634, 2T0S263821, 2T0S263836, 2T0S266610, 2T0S266633, 2T0S264058, 2T0S264081, 8T0S226502, 8T0S229547, 8T0S229559, 8T0S229563, 8T0S229570, 8T0S229574, 8T0S229575, 8T0S229582, 8T0S229583, 8T0S229586, 8T0S229590, 8T0S229601, 8T0S233656, 8T0S233657, 2T0S264214, 0T0S251261, 0T0S251286, 0T0S251319, 0T0S251792, 0T0S251811, 0T0S251809, 0T0S251810, 9T0S236793, 0T0S251309, 2T0S263496, 2T0S263531, 0T0S248136, 0T0S248150, 0T0S248148, 0T0S248156, 1T0S260397, 1T0S260419, 1T0S260349, 1T0S260427, 1T0S260346, 1T0S260433, 1T0S260352, 1T0S260358, 1T0S260370, 1T0S260429, 1T0S260374, 1T0S260367, 1T0S260423, 1T0S260616, 1T0S260637, 1T0S260707, 1T0S260823, 1T0S260405, 1T0S260424, 1T0S260311.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- 2008T BlueStar Hemodialysis Machine 191126Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system5 recalls10 deaths441 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K150708, K222952, K173972 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- 2008T BlueStar Hemodialysis Machine 191130Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls7 deaths272 reports
- 2008T BlueStar Hemodialysis Machine 191124Fresenius Medical Care Holdings, Inc. · Dialyzer, high permeability with or without sealed dialysate system4 recalls32 reports
- 2008® Series Hemodialysis Machines 190828Fresenius Usa, Inc. · Hemodialysis system for home use2 recalls8 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92788).
- Z-0028-2024 · Open, Classified2008T Hemodialysis SYS, with CDX
- Z-0029-2024 · Open, Classified2008T Hemodialysis System without CDX
- Z-0030-2024 · Open, Classified2008T Hemodialysis System w/Bibag
- Z-0031-2024 · Open, Classified2008T GEN 2 Bibag without CDX
- Z-0032-2024 · Open, Classified2008T HD SYS. CDX BLUESTAR
- Z-0034-2024 · Open, Classified2008T HD SYS. W/O CDX BLUESTAR
- Z-0035-2024 · Open, Classified2008T HD SYS. W/O CDX W/BIBAG BLUESTAR
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page