Skip to content

Monoject recallZ-0112-2024 · Cardinal Health 200, LLC · initiated Sep 13, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Syringe contains conical tip that is not compatible with certain needleless IV connector(s).

FDA root cause category: Process change control

What the company is doing

Cardinal Health notified consignees on about 09/13/2023 via letter titled "URGENT MEDICAL DEVICE PRODUCT CORRECTION." Customers were instructed to review inventory for the affected product code and lot, ensure all personnel are informed cease utilizing the affected product with needleless IV connectors, post a copy of the notification in the storeroom where the product is stored, and notify any customers if the product has been further distributed/forwarded. Additionally, customers were asked to complete and return the included acknowledgment form.

Recall details

Recall number
Z-0112-2024
Status
Open, Classified
Initiated by the company
Sep 13, 2023
Posted by the FDA
Oct 13, 2023
Recalling firm
Cardinal Health 200, LLC
Quantity
3.927,994 units
Product code
FMF
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY

Product description

Monoject(TM) 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack, Model Number/Product Code: 1180100777

Code information (models, lots, serial numbers)

Model Number/Product Code: 1180100777. UDI/DI: 10192253034530 each, 20192253034537 box, 50192253034538 case. Lot Numbers: 221201, 221202, 221203, 230201, 230202, 230203, 230204, 230205, 230601.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates