Skip to content

HydroPICC Dual Lumen recallZ-0115-2024 · Access Vascular, Inc · initiated Aug 31, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Access Vascular, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Specific lots of the Dual Lumen HydroPICC Kits were provided with the incorrect revision of the Instructions For Use (IFU) which do not include the following warnings statement: HydroPICC Dual Lumen Catheter is for adult use only. There is a potential risk to health should the HydroPICC Dual Lumen Catheter be used in an inadequately sized vessel. There have been no reported adverse events to date.

FDA root cause category: Error in labeling

What the company is doing

An Urgent Medical Device Correction notification letter dated 8/25/23 was sent to customers. Our records indicate that your facility has received kits from the affected lots. We are contacting you to provide information on action to take to ensure patient safety. - Product that has already been used: No action required. All catheters to date have been used with adult patients. - Product that has not been used: The kit does not have the current IFU. Please refer to the Rev C. IFU provided with this letter. Also, the Patient Information Card, part number LS-067, includes a barcode link to the current revision of the IFU on AVIs website. The following page identifies the shipments you have received with these impacted lots. Please forward this notice in the event the affected product was further distributed. We will contact you to identify the number of kits that are currently in your inventory and determine the appropriate correction with your input, in a timely manner. If you have any questions about this communication, please contact us at 781-538-6594.

Recall details

Recall number
Z-0115-2024
Status
Open, Classified
Initiated by the company
Aug 31, 2023
Posted by the FDA
Oct 13, 2023
Recalling firm
Access Vascular, Inc
Quantity
58 total
Product code
LJS
FDA submissions
K213550
Distribution
US Nationwide distribution in the states of IL, TX.

Product description

HydroPICC 5Fr Dual Lumen Maximal Barrier Kit, Product Model Number 80002004

Code information (models, lots, serial numbers)

UDI-DI: 00850030354068; Lot Number: 11469666

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K213550 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 93043).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page