The 4Kscore Test, an in vitro serum or plasma test that… recallZ-0516-2024 · BioReference Health, LLC · initiated Oct 31, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioReference Health, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Mathematical modeling of the 4Kscore with and without DRE information was recently performed on over 3,000 specimens. DRE information is a required component of The 4Kscore Test.
FDA root cause category: Device Design
What the company is doing
A notification letter dated 10/31/23 was sent to customers. As of the FDA approval of 4Kscore in December 2021, only results with a digital rectal examination (DRE) input of Nodule or No Nodule are considered as valid. Customers who received 4Kscore results without having DRE information inputted (e.g., no information entered for Nodule or No Nodule ) are advised that these results may have increased uncertainty, potentially resulting in erroneous decision interpretations, as DRE information is required for the interpretation recommendations contained in the PMA approved device labeling. The device has now been modified such that if the required DRE information is not provided (e.g., no information entered for Nodule or No Nodule ), then the blood sample will be frozen and stored at our laboratory for up to 60 days, and 4Kscore results will not be performed unless the required DRE information is provided within 60 days. As per the package insert, the DRE cannot be performed within 96 hours before sample collection, but it can be performed after phlebotomy, or in the preceding 12 months before testing. If it is more convenient to schedule the DRE in an office visit after the blood draw, blood samples will be frozen and stored at our laboratory for up to 60 days. The required DRE information can be provided to our customer service center within that me, and the sample will be tested once this information is received. No results will be reported if the input of Nodule or No Nodule information is not received by 60 days. If there are any questions, please feel free to reach out by telephone: 1-833-4KSCORE (1-883-457-2673), Fax: 201-663-4113, or email: [email protected]. Thank you again for allowing BioReference Health to assist you in the care of your patients with potential prostate cancer. Please respond using the page following this letter, by fax 201-663-4113, or email: [email protected] to confirm receipt of thi
Recall details
- Recall number
- Z-0516-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 31, 2023
- Posted by the FDA
- Dec 8, 2023
- Recalling firm
- BioReference Health, LLC
- Quantity
- 1 system (3,196 tests)
- Product code
- QRF
- FDA submissions
- P190022
- Distribution
- US Nationwide distribution.
Product description
The 4Kscore Test, an in vitro serum or plasma test that combines the results of four immunoassays (Roche Elecsys total PSA (prostate specific antigen), Roche Elecsys free PSA, intact PSA, and human kallikrein 2) into a single numerical score that also incorporates the following information: a patient s age, previous biopsy, and digital rectal exam (DRE).
Code information (models, lots, serial numbers)
UDI-DI: 00850038860004
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- 4Kscore Test 1Bioreference Health, LLC · Multi-Analyte Test System With Algorithmic Analysis For Detection Of Prostate Cancer2 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.