Skip to content

FT1000S FLUID TRANSFER SET, REF No. 356048. For direct… recallZ-0128-2024 · B. Braun Medical, Inc. · initiated Sep 26, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from B. Braun Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose both patients and healthcare workers to hazardous medication. There is also the potential for life-threatening situations to occur requiring additional medical intervention.

FDA root cause category: Process control

What the company is doing

On September 26, 2023, the firm notified customers via Urgent Product Recall Notification letters. Customers were informed of the potential for leakage at the male Luer lock site. Customers were instructed to report total amount of impacted product on hand to the firm, after which B Braun Customer Support will contact the customer with instructions on how to return the product for credit or replacement depending on the customer's need. Questions about this recall may be reported to BBMI s Postmarket Surveillance Department by calling 1-833-425-1464.

Recall details

Recall number
Z-0128-2024
Status
Open, Classified
Initiated by the company
Sep 26, 2023
Posted by the FDA
Nov 3, 2023
Recalling firm
B. Braun Medical, Inc.
Quantity
24,400 units
Product code
FPA
FDA submissions
K780551
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and Singapore.

Product description

FT1000S FLUID TRANSFER SET, REF No. 356048. For direct transfer of fluids between large volume parenteral solution containers.

Code information (models, lots, serial numbers)

UDI-DI: 4022495769558 Lots 0061755752, 0061761812, 0061766279

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 93074).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page