Monoject recallZ-0145-2024 · Cardinal Health 200, LLC · initiated Sep 20, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Demonstrated recognition and compatibility issues with syringe infusion pumps.
FDA root cause category: Vendor change control
What the company is doing
Cardinal Health notified customers on about 09/20/2023 via "URGENT MEDICAL DEVICE PRODUCT CORRECTION" letter sent FedEx next day delivery. Customers were instructed to: 1. REVIEW your inventory for the affected product codes and lots. Location of product code and lot are shown in above listed table and below example labeling. 2. COMMUNICATE with all personnel that utilize the Cardinal Health Monoject" Luer-Lock Tip syringes (1, 6, 12, 20, 35 and 60 mL) that they should not be used with syringe infusion pumps. 3. POST a copy of this notification in your storeroom where the product is stored. 4. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this medical device product correction and share a copy of this notice. 5. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMB [email protected], whether you have affected product or not. complete and return the provided acknowledgement form. For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.
Recall details
- Recall number
- Z-0145-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 20, 2023
- Posted by the FDA
- Oct 26, 2023
- Recalling firm
- Cardinal Health 200, LLC
- Quantity
- 3,113,846 units
- Product code
- FMF
- Distribution
- US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
Product description
Monoject 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack
Code information (models, lots, serial numbers)
Product Code: 1180100777; UDI/DI: 10192253034530 - each, 20192253034537 - box, 50192253034538 - case; Lot Numbers: 221201, 221202, 221203, 230201, 230202, 230203, 230204, 230205, 230601
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Monoject 1180100777Cardinal Health 200, LLC · Syringe, piston3 recalls4 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93075).
- Z-0146-2024 · Open, ClassifiedMonoject" 6 mL Syringe Luer-Lock Tip Soft Pack
- Z-0147-2024 · Open, ClassifiedMonoject" 12 mL Syringe Luer-Lock Tip Soft Pack
- Z-0148-2024 · Open, ClassifiedMonoject 20 mL Syringe Luer-Lock Tip Soft Pack
- Z-0149-2024 · Open, ClassifiedMonoject" 35 mL Syringe Luer-Lock Tip Soft Pack
- Z-0150-2024 · Open, ClassifiedMonoject" 60 mL Syringe Luer-Lock Tip Soft Pack
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page