Cvs Pcsu Sjh, Ref Sjcv48j recallZ-0175-2024 · American Contract Systems, Inc. · initiated Sep 1, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
FDA root cause category: Process control
What the company is doing
Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 09/01/2023 by email. The notice explained the issue, risk, and requested the user discontinue the use of the I.V. Cath and Adhesive Dermabond and affix a warning label to each kit and case of affected product in inventory. The firm sent an updated letter on 09/22/2023 via email which corrected the product distribution date.
Recall details
- Recall number
- Z-0175-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 1, 2023
- Posted by the FDA
- Oct 26, 2023
- Recalling firm
- American Contract Systems, Inc.
- Quantity
- 36 cases
- Product code
- LRO
- Distribution
- US Nationwide distribution in the states of FL, IA.
Product description
CVS PCSU SJH, REF SJCV48J
Code information (models, lots, serial numbers)
UDI/DI 00191072167903, Lot Numbers: 945231, 954231, 974231
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ACS SJCV48JAmerican Contract Systems, Inc. · General surgery tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93086).
- Z-0171-2024 · Open, ClassifiedGeneral Pack, REF BBGP31C
- Z-0172-2024 · Open, ClassifiedLAVH, REF BBLV21C
- Z-0173-2024 · Open, ClassifiedBreast Plastics Pack, REF BBPL20D
- Z-0174-2024 · Open, ClassifiedRobotics Pack SAH, REF SAR530M
- Z-0176-2024 · Open, ClassifiedPediatric Heart SJH, REF SJPH34K
- Z-0177-2024 · Open, ClassifiedVasc. Open CVOR SJH, REF SJVO56B
- Z-0178-2024 · Open, ClassifiedRobotics Pack, REF SJVO56B
- Z-0179-2024 · Open, ClassifiedRobotics Pack, REF SSRO22D
- Z-0180-2024 · Open, ClassifiedSub Q Recorder Pack, REF BPRC87A
- Z-0181-2024 · Open, ClassifiedPediatric Urology Wolson Pack, REF BPPU55A
- Z-0182-2024 · Open, ClassifiedCath Cardiac Pack, REF IHCC03V
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page