Skip to content

Senhance Surgical System, which includes: Manipulator Arm… recallZ-0170-2024 · Asensus Surgical Inc · initiated Sep 15, 2023

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Asensus Surgical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.

FDA root cause category: Software design

What the company is doing

Asensus Surgical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 09/15/2023 by mail. The notice explained the problem, risk to health, and requested users to cease use of the device until software version 2.7.5 is installed.

Recall details

Recall number
Z-0170-2024
Status
Terminated
Initiated by the company
Sep 15, 2023
Posted by the FDA
Oct 30, 2023
Recalling firm
Asensus Surgical Inc
Quantity
21 units
Product code
NAY
FDA submissions
K223095
Distribution
US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania

Product description

Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system

Code information (models, lots, serial numbers)

UDI/DI 00815440022075, Serial Numbers: SN 00012, SN 00016, SN 00007, SN 00017, SN 00035, SN 00018, SN 00042, SN 00008, SN 00031, SN 00025, SN 00011, SN 00028, SN 00038, SN 00029, SN 00036, SN 00025, SN 00024, SN 00006, SN 00019, SN 00040, SN 00041, loaded with Software (SW) version 2.7.4

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page