Skip to content

Abbott NT2000iX Radiofrequency (RF) generator, REF… recallZ-0209-2024 · Abbott · initiated Sep 28, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.

FDA root cause category: Process change control

What the company is doing

Abbott issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/28/2023 via FedEx 2nd day delivery. The notice explained the problem, risk, and requested return of the device for inspection. Those devices which pass, will be returned. Should your device(s) fail the inspection, your local Abbott Representative can advise the consignee of alternate commercially available products.

Recall details

Recall number
Z-0209-2024
Status
Open, Classified
Initiated by the company
Sep 28, 2023
Posted by the FDA
Nov 2, 2023
Recalling firm
Abbott
Quantity
34 devices
Product code
GXD
FDA submissions
K111576
Distribution
US, Taiwan, and Colombia

Product description

Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator

Code information (models, lots, serial numbers)

UDI/DI 05415067022417, serial numbers: 16930509, 16983330, 17583407, 17634837, 17783928, 17783930, 17905961, 17928413, 17954521, 17991999, 18047384, 18493711, 18522937, 18675068, 18690920, 18709734, 18742997, 2475-12, 2504-12, 2547-13, 2885-14, 3003-14, 3278-15, 3450-15, 3550-15, 17942357, 18003310, 18102335, 18132017, 18132092, 18657521, 18666647, 19182119, 19197174

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page