StealthStation S8 Application version 2.0 and 2.0.1 (Part… recallZ-0188-2024 · Medtronic Navigation, Inc. · initiated Sep 27, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Navigation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a software anomaly which potentially could result in the surgical planning data shifting to an unintended location.
FDA root cause category: Software design
What the company is doing
On 09/27/2023, the firm sent an "URGENT: MEDICAL DEVICE CORRETION" Letter/Communication via UPS to customer informing them of a software anomaly affecting certain StealthStation S8 and StealthStation FlexENT system with the StealthStation S8 App software version 2.0 and 2.0.1. Due to the software anomaly, there is the potential for the surgical planning data to encounter a displayed shift which could result in proceeding to an incorrect location from the planned target. This result in unintended tissue damage leading to permanent neurological injury, additional pass of device (biopsy needle or electrode), prolonged procedure or need for additional surgery. Customer are instructed to: 1. Follow the below instructions and refer to Appendix A for full details on impacted systems, issues and mitigations: a. Do not change the reference exam in Cranial or ENT procedures, if the surgical planning data is defined on a reference exam merged with pre-merge type or diffusion series. 2. Please review this information, including the additional details in Appendix A, with all physician users. If you have any questions related to this issue, please contact Medtronic Technical Services for help at 1-888-826-5603 or email at [email protected] 3. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at [email protected]. a. Medtronic will contact your facility to coordinate the software update when it is available. 4. This notice needs to be passed on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Questions or assistance, contact Medtronic Sales Representatives or Technical Services at 1-888-826-5603.
Recall details
- Recall number
- Z-0188-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 27, 2023
- Posted by the FDA
- Nov 10, 2023
- Recalling firm
- Medtronic Navigation, Inc.
- Quantity
- 5,240 affected systems
- Product code
- HAW
- FDA submissions
- K212397
- Distribution
- Worldwide Distribution: U.S. (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.(foreign) including countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, FRENCH POLYNES, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PANAMA, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIAN FEDERATION, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UKRAINE, and UNITED KINGDOM
Product description
StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with StealthStation Cranial software, is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. The StealthStation FlexENT System, with StealthStation ENT software, is intended as an aid for locating anatomical structures in either open or percutaneous ENT procedures.
Code information (models, lots, serial numbers)
Model Number: 9735762 UDI-DI Code: 00763000532123 Software Versions 2.0 and 2.0.1 All systems with the impacted software version regardless of how it was purchased Updated list of Product Models/Serial provide on 06/06/2024 and was uploaded to Associated Documents.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- SteathStation 9735762Medtronic Navigation, Inc. · Orthopedic stereotaxic instrument1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page