Skip to content

Regard CV PK, Item Numbers a) 880289014, b) 880289015 recallZ-0315-2024 · ROi CPS LLC · initiated Sep 21, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ROi CPS LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Specific Regard kits contain the Gentige Chest Drain Products being recalled by Gentige. The Instructions for Use (IFU) for the chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains.

FDA root cause category: Other

What the company is doing

ROi CPS, LLC issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to its sole consignee on 09/29/2023 via email. The email attached the notice and the notice received from Getinge. The notice explained the issue, the potential risks, and requested the distributor notify their customer. End user was directed to post the Getinge notice in area where the affect kits were used. An updated notice was sent on 10/13/2023 via email with the corrected item numbers with the same instructions.

Recall details

Recall number
Z-0315-2024
Status
Open, Classified
Initiated by the company
Sep 21, 2023
Posted by the FDA
Nov 16, 2023
Recalling firm
ROi CPS LLC
Quantity
134 units
Product code
OEZ
Distribution
MO

Product description

Regard CV PK, Item Numbers a) 880289014, b) 880289015; cardiovascular convenience kit

Code information (models, lots, serial numbers)

a) UDI/DI 10194717111702, Lot Numbers: 94114, 95436; b) UDI/DI 10194717117049, Lot Numbers: 96442

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page